Calibration
Press
Cal
key 3 seconds and
set Flow Rate calibration
1
Before starting the procedure
ensure that:
•There are no air bubbles in the
pipe
•The graduated tank is empty
• The faucet is off
2
Press
Cal
Key
3
The instrument will show the
following screen
Calibration
Wait________0__
4
The number on the bottom right
indicates the number of pulses
generated by the sensor.
5
Turn on the faucet and fill the
graduated tank up to the required
quantity of water to carry out the
calibration, then turn off the
faucet.
6
After checking that the number of
pulses has completely stopped,
press
Cal
(If the Cal key is
pressed with the number of pulses
set at zero, an error is shown.)
7
Set in the instrument the number
of liters read on the tank
Calibration
10.00 L
8
press
Cal
to confirm
9
The instrument shows the
K
Factor
calculated.
Calibration
K Factor 2.5
10
Press
Cal
to confirm the new
K
Factor
11
Normal Status
enter
cal
enter
cal
enter
cal
enter
cal
enter
cal