The reprocessing of the accessories:
● Breathing circuit:
1.Dismantle all components except the stainless steel spring
attached to the body of water trap.
2.Clean the parts with low flow running water, using a soft
brush to remove all foreign material. (for example, soil or
organic matter)
3.Sterilization by autoclave: 132℃/ 4minutes.
● Chamber:
1. Clean the parts with low flow running water, using a soft
brush to remove all foreign material. (for example, soil or
organic matter)
2. Sterilization by autoclave: 132℃/ 4minutes.
-5-