8.5 The heated wire adaptor disconnect or malfunction, the LED
will flash and audible alarm.
9.
Troubleshooting
Alarm Priority: MID
-20-
ALARM CONDITIONS
REQUIRED ACTION
The alarm tolerance of Airway Temp
is 2℃ above or 4℃ lower than the
temperature control setting, The
Airway Temp LED will flash and
audible alarm.
Please check the
temperature probe and
the water in the circuits.
The alarm tolerance of Chamber
Temp. is above or lower than 4℃
the temperature control setting, The
Chamber Temp. LED will flash and
audible alarm.
Please check the
aluminum tray and
temperature probe.
The Temperature probe disconnect
or malfunction, the LED will flash
and audible alarm.
Please confirm that the
tubing is installed
correctly.
The heated wire adaptor disconnect
or malfunction, the LED will flash
and audible alarm.
Please confirm that the
tubing is installed
correctly.