2. Specification and Definitions
Dimension: 17.5cmX13.5cmX14cm
Weight: 1500g
I/P: 100-120 Vac, 50/60Hz Sine Wave, 3.4A or
200-240 Vac, 50/60Hz Sine Wave, 1.7A
(Please refer to the mark on device.)
Fuse Rating: T4AL/250Vac
Heated Wire Power Supply: DC24V / 120 watts Max
Overheating Protection:
While heated temperature is over 120℃±7℃, the device will cut
out the power automatically.
The VH-3000 life cycle is 3 years.
Operating Environment Temperature:
18~26℃, Humidity RH:15%~95%, Pressure: 700 to 1060hPa
Storage Temperature: 25~40℃, Humidity RH:15%~95%
without condensation
Transport Temperature: -10~60℃, Humidity RH:15%~95%
without condensation, Pressure: 595 to 1013 hPa
Warm-up Time: ≦ 30 minutes
Suggested Flow Range: 2~60 LPM
Humidity Output :>33mg/l within recommended flow range
Standard Compliance: IEC/EN/ES 60601-1, IEC/EN 60601-1-2,
IEC 60601-1-8, ISO 80601-2-74
Alarm Sound Pressure: > 45dB
Classified as: Class I, Type BF, IPX 1
-2-