Assemble the device and housing sections
Before restarting the pumping unit, all parts of the device and
housing which had been removed must be fixed back in place.
1
2
1.
Carefully press the dia-
phragms centrally into the
housing opening, ensuring
they are flush with it.
2.
Hold the pump head pair at the
vacuum pump and wind in the
screw fittings; hex key size 5.
3
4
3.
Tighten the screw connec-
tions crosswise using a torque
wrench with 12 Nm.
4.
Slide the molded hoses back
onto the hose nozzles.