Installation and connection
4.3.3 Coolant connection at the condenser
A vapor condenser EK has a connection for coolant liquids. Water
or the liquid from a chiller, for example, are suitable coolants.
In a closed, in‑house coolant circuit, the pressure should be
limited to 3 bar (44 psi).
A cooling water valve may only be installed in the intake; the
coolant drainage must be clear and non-pressurized.
Connect coolant
(i)
(j)
IN
EX
1.
Fix both hose nozzles
(i)
to the condenser as shown using
union nuts
(j)
.
2.
Fix the hoses for the coolant on the condenser as shown:
IN = feed line, EX = outlet.
3.
Fix the vacuum hoses, e. g., with hose clamps.
Example
Coolant connection
on the EK
Coolant connection
IN = Feed line
EX = outlet
IMPORTANT!