20901220_EN_VACUU·PURE 10_V1.6_060722
Error messages
Error
Error
`
Possible cause
3
Remedy
Personnel
No suction
power or
minimal
suction
power.
`
Leak in inlet line or in
apparatus.
3
Check inlet line
and equipment for
leaks.
Operator
`
Inlet line too long
or cross‑section too
small.
3
Use shorter inlet
line with larger
cross‑section.
`
Condensate inside
the vacuum pump.
3
Let vacuum pump
run for a few
minutes with suction
nozzle open or in
regeneration mode.
`
Deposits inside the
vacuum pump.
3
Clean the pump unit
by operating it with
an open inlet or in
regeneration mode.
`
High level of vapor
generated in the
process.
3
Check process
parameter.
Specialist
`
Pump speed reduced
due to excessive
temperature.
3
See warning,
flashing rate 1x.
Resp.
specialist
Key LEDs do
not light up
`
Vacuum pump
switched off.
3
Switch on vacuum
pump at toggle
switch.
Operator
`
Power plug not
correctly plugged in
or pulled out.
3
Check power
connection and
cable.
`
Vacuum pump
defective.
3
Send in vacuum
pump.
Resp.
specialist
Error – Cause –
Remedy
Summary of Contents for VACUU-PURE 10
Page 1: ...Instructions for use OI no 20901220 Original instructions EN VACUU PURE 10 Vacuum pump...
Page 12: ...12 20901220_EN_VACUU PURE 10_V1 6_060722 Introduction...
Page 30: ...30 20901220_EN_VACUU PURE 10_V1 6_060722 Safety information...
Page 84: ...84 20901220_EN_VACUU PURE 10_V1 6_060722...
Page 85: ...85 20901220_EN_VACUU PURE 10_V1 6_060722...
Page 86: ...86 20901220_EN_VACUU PURE 10_V1 6_060722...