45
20901220_EN_VACUU·PURE 10_V1.6_060722
Commissioning (operation)
>
The vacuum pump does not create a vacuum‑tight seal when
it switches off.
>
If necessary, install an in‑line solenoid valve or shut‑off valve
in the inlet line to isolate your application from the vacuum
pump with a vacuum‑tight seal.
5.2.2
5.2.2 Regeneration mode
Regeneration mode
Regeneration mode is used to quickly dry the pump interior after
the application is finished or before it is taken out of service. Here,
ambient air is fed into the interior of the pump and the interior is
dried by the air supply.
The pump does not need to be separated from the application
for regeneration
.
The pump runs at a reduced speed during regeneration.
The air for regeneration mode is drawn in via a filter on the rear
of the vacuum pump. The ambient air is aspirated here.
>
Check the filter regularly for dirt and blockages.
>
Replace dirty or clogged filters
see chapter:
7.3 Air inlet filter on page 68
7.3 Air inlet filter on page 68
Start regeneration mode
5
The vacuum pump runs at a reduced speed and aspirates
ambient air.
5
The inside of the pump is dried.
5
Regeneration mode ends automatically after a period of
one hour.
Start regeneration
mode
IMPORTANT!
Press key
Drying
(regeneration) with
ambient air
Summary of Contents for VACUU-PURE 10
Page 1: ...Instructions for use OI no 20901220 Original instructions EN VACUU PURE 10 Vacuum pump...
Page 12: ...12 20901220_EN_VACUU PURE 10_V1 6_060722 Introduction...
Page 30: ...30 20901220_EN_VACUU PURE 10_V1 6_060722 Safety information...
Page 84: ...84 20901220_EN_VACUU PURE 10_V1 6_060722...
Page 85: ...85 20901220_EN_VACUU PURE 10_V1 6_060722...
Page 86: ...86 20901220_EN_VACUU PURE 10_V1 6_060722...