8.1 Technical data
Product designation
Product names
Chemistry pumping unit series
PC 3010 VARIO select
PC 3016 VARIO select
PC 3012 VARIO select
PC 3012 VARIO select EKP
Technical data
Technical data
Ambient conditions
(US)
Ambient temperature
10–40 °C
50‑104°F
Altitude, max.
2000 m
above sea level
6562 ft
above sea level
Relative humidity
30–85 %, non condensing
Pollution degree
2
Impact energy
5 J
Protection class (IEC 60529)
IP 40
Type of protection(UL 50E)
Type 1
Prevent condensation or contamination from dust, liquids, or corrosive
gases.
Operating conditions
(US)
Working temperature
10‑40 °C
50‑104 °F
Storage/transport tempera-
ture
-10‑60 °C
14‑140 °F
Maximum admissible media temperature (gas), non-explosive atmos-
phere:
Short term
80 °C
176 °F
Continuous operation
45 °C
113 °F
ATEX conformity
II 3/- G IIC T3 X internal atm. only
Maximum admissible media temperature (gas),
atmosphere:
Short term
40 °C
104 °F
Continuous operation
40 °C
104 °F
Connections
Vacuum, inlet IN
Small flange KF DN 25
Hose nozzle SW 15
Gas ballast GB
Gas ballast valve, manual
Inert gas adapter – OPTION
Small flange GB NT KF DN 16
Hose nozzle GB NT DN 6‑10