Do not pump:
-
unstable substances
- substances which react explosively under
impact
(mechanical stress) without air
- substances which react explosively when being ex-
posed to
elevated temperatures
without air,
- substances subject to
auto-ignition
,
- substances which are
inflammable without air
-
explosive substances
.
➨
The pumps are
not approved
for operation below
ground.
Do not operate
the pump
below ground
.
•
Do not pump
substances which may form
depos-
its
inside the pump. The pumps are not suitable for
pumping substances which may form deposits inside
the pump. Deposits and condensate in the pump may
lead to increased temperatures even to the point of
exceeding the maximum permitted temperatures.
• Check the inlet and outlet of the pump, if there is a
danger of forming
deposits
inside the pump, e.g., in
the pump chambers (the pump chamber is the part
between diaphragm and head cover. See section “Re-
placing diaphragms and valves”, pg. 51). Inspect the
pump chambers regularly and clean if necessary.
•
Consider interactions and chemical reactions of
the pumped media.
Ensure that the materials of the
pump’s wetted parts are compatible with the pumped
substances, see section “Technical data”, pg. 24.
When changing the substances pumped, we recom-
mend purging the pump with air or inert gas prior to
changing the pumped media. Purging the pump will
pump out residues and it will reduce the possibility of
reactions of the pumped substances with each other
and with the pump’s materials.
Summary of Contents for ME 16C NT VARIO o.C.
Page 54: ...page 54 of 96 3 2 4 4 4x...
Page 56: ...page 56 of 96 2 5 mm Emission condenser Peltronic 8 6A Emission condenser Peltronic 4 2x 7A...
Page 57: ...page 57 of 96 Emission condenser Peltronic 4 4x 9 8A 10 4 2x...
Page 58: ...page 58 of 96 12 11 4 4x Don t loosen this screw 12A 13...
Page 59: ...page 59 of 96 5 2x 12A 1 2 3 1 4 1 4 3 2x 5 13 14...
Page 60: ...page 60 of 96 17 5 2x 16 15...
Page 62: ...page 62 of 96 18 2 5 mm Replacing the diaphragms 19 20 Clean 5 8x 20696867 20696839...
Page 63: ...page 63 of 96 21 22 23 1 2...
Page 64: ...page 64 of 96 24 3 25 Pay attention to number of washers 26...
Page 67: ...page 67 of 96 33 34 35 Clean...
Page 68: ...page 68 of 96 36 37 38 IN IN EX EX 4x 20696867 20696839...
Page 69: ...page 69 of 96 39 40 41 IN EX T20 4x 2 2 ft lbf 3 Nm...
Page 70: ...page 70 of 96 42 43 44 Maintain all pump heads in the same way...
Page 74: ...page 74 of 96 2 size 2 2x 3 4A Version A valve 24 4 mm...
Page 75: ...page 75 of 96 5A 20696839 24 4 mm 26 mm 4B 5B 20696839 37 mm 40 mm Version B valve 37 mm...
Page 78: ...page 78 of 96 Assembling the pump housing 46 5 2x 8 9 ft lbf 12 Nm 47 48...
Page 79: ...page 79 of 96 49 5 2x 8 9 ft lbf 12 Nm 50 51...
Page 80: ...page 80 of 96 52 52A 53 52A 1 2 3 3 2x...
Page 81: ...page 81 of 96 53 4 4x 3 7 ft lbf 5 Nm 54 4 2x 3 7 ft lbf 5 Nm 55 4 4x 3 7 ft lbf 5 Nm...
Page 82: ...page 82 of 96 56 57 58...
Page 84: ...page 84 of 96 59 60 4 4x 3 7 ft lbf 5 Nm 61...
Page 87: ...page 87 of 96 3 4 20638821 20638821 5...