page 48 of 96
Troubleshooting
Fault
Possible cause
Remedy
❑
Pump does not
start or stops im-
mediately.
➨
Electrical power cord
not plugged in, electrical
supply failure?
✔
Plug in power cord.
Check fuse.
➨
Device fuse blown?
✔
Identify cause of failure.
Replace device fuse.
➨
Overpressure in outlet
line?
✔
Remove blockage in line,
open valve.
➨
Motor overloaded?
✔
Allow motor to cool down,
identify and eliminate
cause of failure. Manual
reset is necessary. Switch
off pump or unplug.
❑
Pump does not
achieve its ulti-
mate vacuum or
usual pumping
speed.
➨
Centring ring at small
flange connection not
correctly positioned, or
leak in the pipeline or
vacuum system?
✔
Check pump directly -
connect vacuum gauge
directly at pump inlet -
then check connection,
pipeline and vacuum
system if necessary.
➨
Long, narrow vacuum
line?
✔
Use lines with larger di-
ameter, length as short as
possible.
➨
Pump has been exposed
to condensate?
✔
Allow pump to run for
some minutes with atmo-
spheric pressure at the
inlet to purge.
➨
Deposits have been
formed inside the pump?
✔
Clean and inspect the
pump heads.
➨
Diaphragms or valves
damaged?
✔
Replace diaphragms and/
or valves.
➨
Outgassing substances
or vapor generated in
the process?
✔
Check process parame-
ters.
➨
Pump too hot?
✔
Allow pump to cool down.
Determine and eliminate
the cause of overheating.
❑
Pump too noisy.
➨
Atmospheric or high
pressure at the pump
inlet?
✔
Connect hose or silencer
to pump outlet. Be careful
not to cause outlet over-
pressure, especially with
condensable vapors.
Summary of Contents for ME 16C NT VARIO o.C.
Page 54: ...page 54 of 96 3 2 4 4 4x...
Page 56: ...page 56 of 96 2 5 mm Emission condenser Peltronic 8 6A Emission condenser Peltronic 4 2x 7A...
Page 57: ...page 57 of 96 Emission condenser Peltronic 4 4x 9 8A 10 4 2x...
Page 58: ...page 58 of 96 12 11 4 4x Don t loosen this screw 12A 13...
Page 59: ...page 59 of 96 5 2x 12A 1 2 3 1 4 1 4 3 2x 5 13 14...
Page 60: ...page 60 of 96 17 5 2x 16 15...
Page 62: ...page 62 of 96 18 2 5 mm Replacing the diaphragms 19 20 Clean 5 8x 20696867 20696839...
Page 63: ...page 63 of 96 21 22 23 1 2...
Page 64: ...page 64 of 96 24 3 25 Pay attention to number of washers 26...
Page 67: ...page 67 of 96 33 34 35 Clean...
Page 68: ...page 68 of 96 36 37 38 IN IN EX EX 4x 20696867 20696839...
Page 69: ...page 69 of 96 39 40 41 IN EX T20 4x 2 2 ft lbf 3 Nm...
Page 70: ...page 70 of 96 42 43 44 Maintain all pump heads in the same way...
Page 74: ...page 74 of 96 2 size 2 2x 3 4A Version A valve 24 4 mm...
Page 75: ...page 75 of 96 5A 20696839 24 4 mm 26 mm 4B 5B 20696839 37 mm 40 mm Version B valve 37 mm...
Page 78: ...page 78 of 96 Assembling the pump housing 46 5 2x 8 9 ft lbf 12 Nm 47 48...
Page 79: ...page 79 of 96 49 5 2x 8 9 ft lbf 12 Nm 50 51...
Page 80: ...page 80 of 96 52 52A 53 52A 1 2 3 3 2x...
Page 81: ...page 81 of 96 53 4 4x 3 7 ft lbf 5 Nm 54 4 2x 3 7 ft lbf 5 Nm 55 4 4x 3 7 ft lbf 5 Nm...
Page 82: ...page 82 of 96 56 57 58...
Page 84: ...page 84 of 96 59 60 4 4x 3 7 ft lbf 5 Nm 61...
Page 87: ...page 87 of 96 3 4 20638821 20638821 5...