page 32 of 70
Prevent internal condensation, transfer of liq-
uids or dust. The diaphragms and valves will
be damaged, if liquids are pumped in signifi-
cant amounts.
Check the pump regularly for external soiling
and deposits. Clean the pump if necessary
to avoid an increase of the pump’s operating
temperature.
In case of overload, the motor is shut down by
a
self-hold thermal cutout
in the winding.
Note
: Only manual reset is possible. Switch
off the pump and disconnect from the power
source. Identify and eliminate the cause of fail-
ure. Allow the pump to cool down sufficiently
before restart.
°C / °F
I
➨
O
!
3.
2.
1.
IN
✘
Avoid overheating (e.g., due to hot process
gases). Maximum permissible temperature
range: see “Gas inlet temperatures”, pg. 24.
Do not start the pump if the
pressure difference between
inlet and outlet ports exceeds max. 16.0 psi (1.1 bar)
.
Attempts to start the pump at higher pressure difference
may cause stalling and damage of the motor.
NOTICE
Make sure the fan’s air supply is adequate. Do
not place the pump on soft surfaces (e.g., rub-
ber foam) during operation. This may reduce
the air circulation in the pump housing and af-
fect the fan’s cooling effect. Check fan regu-
larly for dust/dirt. Clean fan guard grill if neces-
sary to avoid a reduction of ventilation.
air
NOTICE
IN
✘
condensate / dust
NOTICE
NOTICE
NOTICE
Summary of Contents for ME 16 NT
Page 42: ...page 42 of 70 3 4 5 4 4x 4 4x 4 2x...
Page 43: ...page 43 of 70 Don t loosen this screw 7 8 6 5 2x...
Page 44: ...page 44 of 70 9 10 11 5 2x...
Page 46: ...page 46 of 70 12 2 5 mm 13 14 2 5 mm 5 8x...
Page 47: ...page 47 of 70 15 696819 696827 16 17 Clean...
Page 48: ...page 48 of 70 696819 696827 18 19...
Page 50: ...page 50 of 70 Assembling the pump head 23 24 25 Pay attention to number of washers...
Page 52: ...page 52 of 70 29 30 31...
Page 54: ...page 54 of 70 35 Maintain all pump heads in the same way ME 16 NT MD 12 NT MV 10 NT 36A 36B...
Page 56: ...page 56 of 70 4 5 6 696827...
Page 57: ...page 57 of 70 7 8 size 2 2x 9 size 2 2x...
Page 59: ...page 59 of 70 Assembling the pump housing 37 38 39 5 2x 8 9 ft lbf 12 Nm...
Page 60: ...page 60 of 70 40 42 41 5 2x 8 9 ft lbf 12 Nm...
Page 61: ...page 61 of 70 45 4 2x 3 7 ft lbf 5 Nm 44 4 4x 3 7 ft lbf 5 Nm 43...
Page 62: ...page 62 of 70 46 4 4x 3 7 ft lbf 5 Nm 47 4 4x 3 7 ft lbf 5 Nm 48...