To check the valves and the diaphragms use a hex key
to remove four socket-head screws from the pump head
and remove the upper part of the housing (housing cover
with housing cover insert) together with the head cover,
the valves and the O-rings.
☞
Never remove parts by using a spiky or sharp-edged tool
(e.g. screw driver), we recommend to use a rubber mallet
or compressed air (to be blown carefully into port).
➨
Remove the head cover carefully from the upper part of
the housing. Note the position of the valves and remove
them.
☞
Replace valves and/or O-rings if damaged.
☞
Use petroleum ether or industrial solvent to remove de-
posits. Do not inhale.
Outlet side
MV 10C VARIO-B / PC 3010 VARIO
MD 12C VARIO-B / PC 3012 VARIO
Inlet side
Pumps VARIO-B / Pumping units PC 301x VARIO