3. - - - - - - - The output signal of Load Cell is inadequate for Span Calibration.
—To decrease the Display Resolution.
–To change Load Cells with higher output signal.
4. U U U U U U U a. The input signal is too high (more than +15 mV)
—Replace new Load Cells with bigger capacities.
b. Load Cells were damaged.
–To replace with new Load Cells.
c. Poor connection.
—To check the connector wirings.