11
¡
¿
cAL.-0
¡
¿
000000
Maximum Capacity: as long as division/full capacity 1/30000, then the full
capacity can be set freely.
Division: it is repeat in the cycle of 1, 2 5, 10, 20, 50. When the maximum capacity
is set, the program can automatically determine the adequate division with as
if the maximum display resolution is 1/30000.
Decimal Place (0, 0.0, 0.00, 0.000 repeat in cycle)
Zero Tracking Range
3.0-2
¡ G
3.0 div. /2 sec.
3.0-1
¡ G
3.0 div. /1 sec.
2.0-2
¡ G
2.0 div. /2 sec.
2.0-1
¡ G
2.0 div. /1 sec.
1.5-2
¡ G
1.5 div. /2 sec.
1.5-1
¡ G
1.5 div. /1 sec.
1.0-2
¡ G
1.0 div. /2 sec.
1.0-1
¡ G
1.0 div. /1 sec.
0.5-2
¡ G
0.5 div. /2 sec.
0.5-1
¡ G
0.5 div. /1 sec.
no.
¡ G
zero tracking disable
Motion Detection Division Range
3d
¡ G
Change within 3 divisions is considered as stable (MOTION Signal off)
1d
¡ G
Change within 1 division is considered as stable (MOTION Signal off)
4.3 CALIBRATION
Press SM when display shows “SEt”, indicator will show “CAL” for a short while before
entering the calibration menu. When all sub-items are set, press SET to enter for the next
selection.
¡ i
ZERO
¡ j
¡ i
SET
¡ j
¡ i
SET
¡ j
dP.
Ct.
C03000
o.tr.
no.d.