WARNING
Do not connect the power cable to a power source until all connections are made. The power
source has to be grounded and protected by a circuit breaker.
WARNING
The UVIDOC HD6 system is designed to fit a specific voltage. Please, check the voltage to ensure it
corresponds to the UVIDOC HD6 specifications.
WARNING
Please keep an open area of at least 20 cm at the rear of the cabinet to ensure a proper air
circulation for the system. The system should be located in an area free of excessive dust or
moisture, strong magnetic fields or ionising radiation. It is also recommended that the ambient
temperature be stable and within the range of 15°C to 25°C (20°C recommended) and that the
relative humidity not exceed 70%, non-condensing.
WARNING
Ensure that all of the systems ventilation openings are free of interference. Excessive heat build-
up in the instrument may effect performance or cause operational failure.
The UVIDOC HD6 system should be located away from water, solvents, or a corrosive material, on
a bench top that is dry and stable. The system should be placed away from interfering electrical
signals and magnetic fields. A dedicated electrical outlet should be used to eliminate electrical
interference from other instrumentation in your laboratory
WARNING
Do not defeat any instrument interlocks; they are designed to prevent user injury.
It is compulsory to power down the system and disconnect the AC mains from the unit before
performing any disassembly or repair to the system.