AUTO

000

A

B

 

Zeroed in AUTO mode

 

 
The 

AUTO

 legend will be blinking and the arrow can be 

pointing to any color. 
The background density of the paper is automatically 
zeroed out, helping the pressman maintain optimum 
appearance across varying paper stocks.  Strongly tinted 
stocks can be zeroed out manually. 
 
Place the aperture on the solid color target to be 
measured.  Press the unit down and hold it while the 
measurement is taken.  The entire process takes less 
than one second. 
 

 

3

Summary of Contents for BETACOLOR S1

Page 1: ...the need of power consuming motors or bulky rechargeable battery packs The AC charger and internal charging circuits have been eliminated Color density measurements are made simultaneously in all four...

Page 2: ...tting Started Open the densitometer by sliding the gray release latch forward allowing the aperture to spring forward Tap either button to turn on the unit if it has powered down Select AUTO mode with...

Page 3: ...the paper is automatically zeroed out helping the pressman maintain optimum appearance across varying paper stocks Strongly tinted stocks can be zeroed out manually Place the aperture on the solid col...

Page 4: ...ll four color channels each time a sample is measured The individual color channel density values can be displayed by stepping through the color channels with button B the right hand button This infor...

Page 5: ...ro all channels Flashing AUTO will display and 0 00 will appear in the readout 3 Measure the solid target DENS will appear the density will be displayed and the color indicator will move to the proper...

Page 6: ...ro function 1 Tap either key to activate the instrument If MAN DENS does not appear at the right of the display press and hold B for three seconds with the head raised MAN DENS 000 A B Yellow channel...

Page 7: ...own The density of the sample as measured in the selected color channel will be displayed The other color component values of that color may be displayed by pressing B after raising the head AUTO DENS...

Page 8: ...DENS and the color indicator will appear along with the measured value 3 Raise the head and press A and B simultaneously ADJ will appear and the current slope value will be displayed Raise or lower t...

Page 9: ...adjusting all four channels as described above 6 Press A and B simultaneously with the head raised to exit the ADJ mode Serial Port Baud Rate Adjustment Adjust the baud rate of the unit to match the r...

Page 10: ...unit and remove the battery cover plate Discard the old batteries and install two new ones in the same positions Reset Button If the unit fails to respond or an unusual display appears press the red...

Page 11: ...e and date stamped data and plots can be saved to disk or printed in color The network client version supports centralized data collection Statistical analysis screen of QC01 BetaTab Data Collection S...

Page 12: ...a Color Proofing Viewer is fitted with a 10x color corrected lens a brilliant tungsten halogen lamp and color separation filters Verify register dot quality coverage and more in every color including...

Page 13: ...8 pin RS232 300 9600 baud Weight 10 4 oz 290 g Dimensions 3 x 5 5 x 1 5in 145 x 75 x 40mm Warranty Your densitometer is warranted to be free of manu facturing defects for two years from the date of pu...

Reviews: