Service Manual
33
Touch Currents
Using an electrical safety analyzer, measure the following conditions. Acceptable conditions are
listed in parentheses:
Ground lead intact
Pass Normal Polarity ______ (<100
µ
A)
Pass Reverse Polarity ______ (<100
µ
A)
Ground lead open
Pass Normal Polarity ______ (<100
µ
A)
Pass Reverse Polarity ______ (<100
µ
A)
Using the electrical safety analyzer, and following the analyzer's instructions, make the
appropriate connections for the following leakage currents. Measure the leakage currents without
activating the system. The active electrode is the only Applied Part subject to these
tests.
Pass Patient Auxiliary Current
_____ (<100
µ
A)
Pass Patient Leakage Current (PLC)
_____ (<100
µ
A)
Pass PLC Caused by Ext Voltage
_____ (<5000
µ
A)
(aka Mains on Applied Part)
CQM Circuit Tests
Do not activate the system during these tests, as most ES analyzers are not designed to
withstand activation during CQM testing.
Following the instructions for the ES analyzer, perform the following measurements IN THE ORDER
LISTED, without disconnecting the CQM dispersive pad lead between steps. Set CQM tester (or
resistor substitution box if analyzer does not have CQM test capability) to 150
Ω.
Connect pad
plug to system and turn on main power switch.
Pass Verify
error is active.
Pass CQM dispersive pad illuminates the CQM pad icon
above the pad receptacle.
Pass Reduce resistance until
error clears.
130
Ω
_____ (117
Ω
-143
Ω
)
Pass Reduce to 50
Ω
, then raise until
error activates.
68
Ω
_____ (61
Ω
-75
Ω
)
Pass Reduce resistance until
error activates.
10
Ω
_____ (5
Ω
-15
Ω
)
Other Error Checks
Pass Pressing both cut and coag switchpen buttons or footswitch pedals simultaneously
produces
error and disables output
Pass Activation of coag mode into open circuit conditions does not produce
error
Summary of Contents for FINESSE +
Page 12: ...8 FINESSE Service Manual ...
Page 18: ...14 FINESSE Service Manual ...
Page 24: ...20 FINESSE Service Manual ...
Page 34: ...30 FINESSE Service Manual ...
Page 42: ...38 FINESSE Service Manual Mains Power and System Wiring Diagram ...
Page 43: ...FINESSE Service Manual 39 ...
Page 45: ...FINESSE Service Manual 41 RF Power Amplifier PCB Drawing ...
Page 46: ...42 FINESSE Service Manual Logic PCB Drawing 90738 B ...
Page 47: ...FINESSE Service Manual 43 Display PCB Drawing 90737 B ...
Page 48: ...44 FINESSE Service Manual Power Supply PCB Drawing 90740 ...
Page 52: ...48 FINESSE Service Manual Notes ...
Page 53: ...FINESSE Service Manual 49 ...