Service Manual
13
The active electrodes, pens, dispersive pads, and speculum tubing/reducer are supplied as single
patient use items. Do not attempt to clean and/or resterilize these items.
Annual Maintenance
CAUTION: Electric Shock Hazard.
Unplug system before replacing internal filter.
The Internal Filter (item no. SSE-500) should be changed on an annual basis.
The
following procedure should be performed using caution appropriate for the handling of
contaminated medical waste:
1. Remove the internal filter retaining ring from the front of the system by rotating it
to the "unlocked" ( ) position, then pull outward. Dispose of the used filter consistent with
your facility's policy.
2. Place the new Internal Filter inside the smoke evacuator port until it is nearly flush
against the front panel.
3. Replace the internal filter retaining ring, lining up the position pointer on the ring with the
"unlocked" icon, inserting the ring until it is flush against the front panel, and rotating until
the position pointer is aligned with the "locked" ( ) icon.
For the care and performance of your electrosurgical system, Utah Medical Products
recommends users establish a program for inspection and preventive maintenance (IPM). Contact
Utah Medical Products at (800) 533-4984 for details regarding IPM service options. UTMD
recommends annual IPM for the system. See the
Test Procedures
section of this
manual.
ECRI, a non-profit organization, publishes
IPM Procedure 411
. This procedure recommends that
minor inspections should be performed every six months and major inspections should be done
annually. Contact ECRI at (610) 825-6000, or www.ecri.org
.
Summary of Contents for FINESSE +
Page 12: ...8 FINESSE Service Manual ...
Page 18: ...14 FINESSE Service Manual ...
Page 24: ...20 FINESSE Service Manual ...
Page 34: ...30 FINESSE Service Manual ...
Page 42: ...38 FINESSE Service Manual Mains Power and System Wiring Diagram ...
Page 43: ...FINESSE Service Manual 39 ...
Page 45: ...FINESSE Service Manual 41 RF Power Amplifier PCB Drawing ...
Page 46: ...42 FINESSE Service Manual Logic PCB Drawing 90738 B ...
Page 47: ...FINESSE Service Manual 43 Display PCB Drawing 90737 B ...
Page 48: ...44 FINESSE Service Manual Power Supply PCB Drawing 90740 ...
Page 52: ...48 FINESSE Service Manual Notes ...
Page 53: ...FINESSE Service Manual 49 ...