Test Procedures
This checklist is intended to provide information for the preventive maintenance (PM) of the
Electrosurgical Generator and Smoke Evacuation Systems. It is intended for use by
technical personnel who have experience with electrosurgical generator operation and
maintenance, and who possess the test equipment and tools necessary to obtain the requested
data. This document may not reference all tests necessary to demonstrate compliance to all
regulations, nor should all tests listed be considered necessary for the safe use of the .
Technical personnel should consult local regulations and industry recommendations to develop
their own protocol for inspection and PM of the .
Required Equipment and Tools
• Electrosurgical analyzer and leads
• Electrical safety analyzer, with patient lead connections
• Two-button electrosurgical switchpen and/or two-pedal footswitch
• SSE-500 Internal Smoke Evacuation Filter
Mechanical Inspection and Service
Pass
Verify front panel controls, switches, and dispersive pad receptacle/pins are not
damaged.
Pass
Verify switchpen receptacles firmly hold switchpen.
Pass
Knobs and color coded knob caps are securely attached.
Pass
Verify power cord and its rear panel receptacle are not damaged.
Pass
Verify footswitch cord (if present) and its rear panel receptacle are not damaged.
Done Replace Internal Smoke Evacuation Filter
Performance Inspection
Pass
Green LED in main power switch lights when switch is toggled to the " " position.
Pass
One and only one cut mode indicator (“C”, “1”, “2”, or “3”) is illuminated..
Pass For all four cut/blend modes, the cut mode waveform selector knob adjusts so that light
indicator for each mode illuminates when knob’s index pointer corresponds to the
correct indicator light.
Verify power displays can be adjusted through the full output range
Pass
Cut: 05-99
Pass Coag: 05-75
Press cut button on switchpen (repeat with cut pedal on footswitch, if available), and verify:
Pass
Yellow cut activity indicator
lights and cut mode tone is audible
Pass
Smoke evacuation motor activates, and remains running 5 seconds after deactivation
Press coag button on switchpen (repeat with coag pedal on footswitch, if available), and verify:
Pass
Blue coag activity indicator
lights and coag mode tone is audible
Pass
Smoke evacuation motor activates, and remains running 5 seconds after deactivation
Summary of Contents for FINESSE +
Page 12: ...8 FINESSE Service Manual ...
Page 18: ...14 FINESSE Service Manual ...
Page 24: ...20 FINESSE Service Manual ...
Page 34: ...30 FINESSE Service Manual ...
Page 42: ...38 FINESSE Service Manual Mains Power and System Wiring Diagram ...
Page 43: ...FINESSE Service Manual 39 ...
Page 45: ...FINESSE Service Manual 41 RF Power Amplifier PCB Drawing ...
Page 46: ...42 FINESSE Service Manual Logic PCB Drawing 90738 B ...
Page 47: ...FINESSE Service Manual 43 Display PCB Drawing 90737 B ...
Page 48: ...44 FINESSE Service Manual Power Supply PCB Drawing 90740 ...
Page 52: ...48 FINESSE Service Manual Notes ...
Page 53: ...FINESSE Service Manual 49 ...