![Utah Medical Products Finesse ESU-110 Service Manual Download Page 25](http://html.mh-extra.com/html/utah-medical-products/finesse-esu-110/finesse-esu-110_service-manual_1064776025.webp)
Waveform Verification
Set and verify the following output waveforms with an isolated oscil-
loscope, into open circuit conditions
Pass
Corrected Frequency ________kHz
(400-500)
Pass
Corrected "CUT"
continuous 450 kHz sinusoid
Pass
Corrected "BLEND 1"
10 cycles on 6 cycles off
Pass
Corrected "BLEND 2"
8 cycles on 8 cycles off
Pass
Corrected "BLEND 3"
6 cycles on 10 cycles off
Pass
Corrected "COAG"
1 cycle on and 15 cycles off
Output Calibration
Check system output power at the following displayed settings. All
measurements should be made through a 500
Ω
load. Tolerances are
listed in parentheses, and trimpot for adjustment in square brackets.
Do not activate system for more than 5 seconds for each reading,
and allow at least 15 seconds between readings.
Pass
Corrected Cut
[R142]
30: _____
60: _____
90: _____
(25-35W)
(51-69W)
(77-103W)
Pass
Corrected Blend 1
[R144]
30: _____
60: _____
90: _____
(25-35W)
(51-69W)
(77-103W)
Pass
Corrected Blend 2
[R146]
30: _____
60: _____
90: _____
(25-35W)
(51-69W)
(77-103W)
Pass
Corrected Blend 3
[R148]
30: _____
60: _____
90: _____
(25-35W)
(51-69W)
(77-103W)
Pass
Corrected Coag [R150]
30: _____
60: _____
75: _____
(25-35W)
(51-69W)
(64-86W)
RF Leakage Currents
Set cut display to “99” and coag display to “75”. Disconnect disper-
sive lead from analyzer and connect the Finesse system chassis
ground (equipotential connector on rear panel) to the dispersive
input on the analyzer. Check RF output through 200
Ω
load:
Pass
Corrected Cut
_____ (<4W)
Pass
Corrected Blend 1
_____ (<4W)
Pass
Corrected Blend 2
_____ (<4W)
Pass
Corrected Blend 3
_____ (<4W)
Pass
Corrected Coag
_____ (<4W)
Chassis Leakage Currents
Using an electrical safety analyzer, measure the following conditions.
Acceptable conditions are listed in parentheses:
Ground lead intact
Pass
Corrected Normal Polarity
______ (<100
μ
A)
Pass
Corrected Reverse Polarity
______ (<100
μ
A)
Ground lead open
Pass
Corrected Normal Polarity
______ (<100
μ
A)
Pass
Corrected Reverse Polarity
______ (<100
μ
A)
Alarm Checks
Pass
Corrected Pressing both cut and coag switchpen buttons
simultaneously produces alarm and disables
output
Pass
Corrected Pressing both cut and coag foot pedals (if
used) simultaneously produces alarm and dis-
ables output
Pass
Corrected A 900 ohm resistor placed between dispersive
pad receptacle pins does not correct patient
alarm condition
Pass
Corrected With jumper in place between U301 pins 1
and 16, activation of any output mode results
in a continuous alarm
Pass
Corrected Activation of coag mode into open circuit
conditions does not produce output error
alarm
Pass
Corrected Each mains lead connection, tested individu-
ally, withstands 1200 VAC relative to chassis
Pass
Corrected Adjust volume control so that cut and coag
audible tones can be clearly heard over the
smoke evacuation motor set at the ‘high’ flow
rate position.
Finesse Service Manual
21
Summary of Contents for Finesse ESU-110
Page 4: ......
Page 10: ...Finesse Service Manual 6 ...
Page 27: ...Mains Power Wiring Diagram Finesse Service Manual 23 ...
Page 28: ...Assembly Diagram Finesse Internal View Finesse Service Manual 24 ...
Page 29: ...RF Power Board Assembly Drawing Finesse Service Manual 25 ...
Page 30: ...Logic Board Assembly Drawing Finesse Service Manual 26 ...
Page 31: ...Display Board Assembly Drawing Power Supply Board Assembly Drawing Finesse Service Manual 27 ...
Page 32: ...System Functional Diagram Finesse Service Manual 28 ...