The negative-feedback output stage limits the output voltage under
high resistance loads. The same feedback circuit attempts to main-
tain the output voltage under low resistance loads so that the cutting
characteristics are quite uniform throughout the entire cut. See
Figure 3
and
Figure 4
for output power vs. load data for cut, blend
and coag modes. Cut mode maximum output power is 150 watts at
200
Ω
patient load.
Output Power vs. Displayed Setting
The output of the Finesse
®
unit increases linearly with adjustment of
the front panel output settings.
Figure 5
and
Figure 6
show the typi-
cal output power of each of the five modes over the full control
range. Both curves shown indicate output at 500
Ω
load resistance.
Output Safety Circuit Specifications
The Finesse
®
system is equipped with a safety circuit that monitors
the output signal levels and disables all unit functions when the out-
put exceeds the expected "nominal" output. The alarm system can
only be reset by turning off the main power switch.
The output safety alarm system will activate if the output voltage
exceeds 180 volts rms or exceeds the nominal output voltage by
50%, whichever is greater.
Audible Tone/Alarm System
Volume:
65 dBA min @ 1m (3.3 ft)
Frequency:
Cut
0.8 kHz nominal
Coag
1.5 kHz nominal
Alarm 3.0 kHz nominal
Smoke Evacuator System
Supply Voltages (all voltages are AC rms)
Nominal Voltage:
115 volts
230 volts
Operating Range:
90-140 volts
180-280 volts
Frequency:
45-65 Hz
45-65 Hz
Maximum Supply Current and Power
At the Nominal Voltage, the maximum supply current is 2 amps.
Maximum supply power for the ESU-110 is 230 watts and for the
ESU-220 is 460 watts.
Overcurrent Protection
The smoke evacuation motor is protected by a 3 amp fuse in the hot
lead.
Air Flow vs. Supply Voltage
At a supply voltage within the Regulated Range, the air flow through
a new disposable filter coupled to a 15 cm long 7 mm ID tube is not
less than 100 liters per minute (3.5 cubic feet per minute) at the
"high" setting and not less than 70 liters per minute (2.5 cfm) at the
"normal" setting.
Smoke Evacuation Duration
The smoke evacuation system will begin running immediately on
activation of the electrosurgery module and remain running five sec-
onds after the output power is deactivated.
Disposable Filter Cartridge
The Finesse
®
Filter Pack, catalog number ESU-501, consists of a
pleated HEPA paper filter followed by a compartment containing
activated charcoal.
Filter Life
The Finesse
®
Filter Pack, to ensure total elimination of odors between
procedures, should be replaced on a daily basis. However, the
three-stage filtration system has been tested to effectively remove
odors and particles for up to 15 electrosurgical procedures. Over a
period of time the external disposable filter pack can be a source of
odor and possible viral contamination. Therefore, it is recommend-
ed that the external filter pack be changed every day or after 15 pro-
cedures if more than 15 procedures are performed in a single day.
Finesse Service Manual
8
Summary of Contents for Finesse ESU-110
Page 4: ......
Page 10: ...Finesse Service Manual 6 ...
Page 27: ...Mains Power Wiring Diagram Finesse Service Manual 23 ...
Page 28: ...Assembly Diagram Finesse Internal View Finesse Service Manual 24 ...
Page 29: ...RF Power Board Assembly Drawing Finesse Service Manual 25 ...
Page 30: ...Logic Board Assembly Drawing Finesse Service Manual 26 ...
Page 31: ...Display Board Assembly Drawing Power Supply Board Assembly Drawing Finesse Service Manual 27 ...
Page 32: ...System Functional Diagram Finesse Service Manual 28 ...