DEFINITION OF TERMS

In the context of this guide, the underlisted words are presumed to mean: 
Division: The amount of increments a scale offers. How accurate the scale can be
Capacity: the maximum amount the scale can contain
Initial Zero Range: The percentage of weight allowed on the scale when indicator is powered 

on that will automatically zero.
example: If initial zero range is set to 10% of the max. capacity and your max. capacity is 

100lbs, you can place up to 10lbs of weight on the scale and when the indicator is powered 

on, it will automatically zero out the weight.
Manual Zero Range: The percentage of weight allowed on the scale where the indicator will 

let you manually zero (anything above this percent will be tared)
Zero Tracking Range: A subset to the manual zero range; if the weight on the scale is not 

stable, the zero tracking range still allows you to zero within a set division of the scale
Zero Tracking Time: A subset to the zero tracking range, it is the time allowed for the scale to 

fall within the zero tracking range tolerance and still qualify to be zero’d
Overload Range: Weight allowance that is out of the set calibrated range. Adds a tolerance to 

the calibrated max. capacity without having to recalibrate.
example: If your scale has a max. capacity of 1000lbs with a division of 1 and you set the 

overload range to 60, you can add 1060lbs of weight to the scale without it displaying an error 

code
Negative Display: How far you can go in the negative direction before displaying an error 

code
Standstill Time: How fast the scale will stabilize
Standstill Range: How much the scale can fluctuate before being determined stable
Digital Filter: For filtering moving weight, such as animals, It changes how sensitive the scale 

is to variations in movement.
Noise Filter: A filter for how susceptible the scale is to general variations
Baud Rate: The rate at which information is transferred in a communication channel. 

example: In the serial port context, “9600 baud” means that the serial port is capable of 

transferring a maximum of 9600 bits per second.

Summary of Contents for US-6011 Series

Page 1: ...US 6011 SERIES INDICATOR USER S MANUAL US 6011SS 4 20MA US 6011SS R usameasurements com...

Page 2: ...4 PREPARATION SET UP 5 POWER SUPPLY 7 DISPLAY AND KEY DESCRIPTION 8 OPERATING INSTRUCTIONS 9 CALIBRATION 12 INDICATOR PARAMETER SETTINGS 14 CONNECTORS 21 RS232 SERIAL OUTPUT FORMAT 23 US 6011SS 4 20 m...

Page 3: ...ubset to the zero tracking range it is the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the...

Page 4: ...ed Maximum capacity to prevent damaging the inner load cell 9 Stay with in the unit s rated load limit 10 Do not leap on the scale 11 Only professional staff should conduct the calibration inspection...

Page 5: ...date Modify gravity based on location 100 manual zero range setting Automatic error warning alarm Splash proof keyboard and display RS 232 Port for optional Scoreboard RS 232 Port for connecting to Pr...

Page 6: ...pn 6 ppm K Sensitivity internal 0 3 v d Input voltage 30 to 30Mv DC Excitation circuit 5 VDC 4 wire connection 6 load cell of 350ohm max AC power AC 100 250V use only the included 9V adapter supplied...

Page 7: ...month to keep the battery in its best condition When you want to discharge the battery you are required to leave the indicator on until the indicator powers goes off Remove the battery when it is not...

Page 8: ...t product ZERO Indicates that the scale is zero PRINT Means data should be printed confirms inputs Shows that the scale is at zero Shows that the scale is stable Gross Shows when in Gross weight mode...

Page 9: ...e indicator is in gross mode When you do that the current weight on the scale will be tared and the scale will enter the net mode Add your product to the scale to weigh without the pallet s weight If...

Page 10: ...ing between 2 sets of numbers If you want to print the accumulated total hold the PRINT key for one second while the last 4 digits of the total are shown To exit accumulation mode wait for the last 4...

Page 11: ...the average weight from those 3 seconds Press HOLD again to exit holding mode Print If the indicator is connected to a printer and the weight on the scale is stable press the PRINT key to print the c...

Page 12: ...C2 set to 1 the scale will read in 0 1 lb increments 12 Press PRINT to set the value The display will now show C04 13 Press PRINT to access the C4 channel The display will show 14 Enter in the maximum...

Page 13: ...tightly and verify that the platform is level To Calibrate using only multiple calibration weights Linear Calibration 1 Press PRINT to continue The display will now show C06 2 Press PRINT to access t...

Page 14: ...g or lb 2 Press and hold the HOLD and PRINT key at the same time for 2 seconds 3 Navigate through the settings C01 to C45 as shown in the table 4 below by using the arrow buttons and return buttons as...

Page 15: ...ge within 3 minutes 5 turn off display if no change within 5 minutes LCD Version OP900B 0 turn off the backlight 1 backlight only when the weight changes or keyboard is pressed2 constant backlight Hol...

Page 16: ...adhesive label thermal printer 7 Print to Zebra large adhesive label thermal printer 8 Reserved Baud Rate C19 0 1200 for OP 910 remote display 1 2400 2 4800 3 9600 for all printers and OP 910X OP 910...

Page 17: ...verload range setting Negative Display C25 0 9d 10 10 max capacity 20 20 max capacity 50 50 max capacity 100 100 max capacity Standstill Time C26 0 quick 1 medium 2 slow Standstill Range C27 d divisio...

Page 18: ...elay output function 1 2 turn on relay output function 2 3 Reserved menu Gravity of Calibration Location C36 9 7000 9 9999 Gravity of Destination C37 9 7000 9 9999 Version No C38 Print Mode C41 0 auto...

Page 19: ...Weighing Unit C01 1 Decimal Setting C02 0 Graduation Setting C03 1 Maximum Capacity C04 1000 Zero Calibration C05 0 Calibration C06 0 Restore Default C07 0 Warning Tone C08 1 Automatic Power Off C09 0...

Page 20: ...ter C29 2 Print Time and Date C30 0 Analog Output Setting C31 1 Calibrate Current C32 4 Relay Output Setting C33 1 Multi connection add C34 0 Wireless Communication C35 6 Wireless Communication C36 9...

Page 21: ...ods between the load cell and indicator Quick Disconnect as shown below Hardwire Using Inner Terminal Block Connection Note Make sure you follow all the anti static rules to avoid damage to your indic...

Page 22: ...Negative output signal from load cell EXC EX BLACK Negative excitation voltage to load cell DB9 Connection 9 pin Serial Connector The DB9 9 pin serial connector is used for different purposes dependin...

Page 23: ...is a list of the serial parameters 8 data bits 1 stop bits No parity No handshaking Note With the RS232 Output option we have data logging software available as seen in Insert figure no Below are the...

Page 24: ...XXXXXXX 1 1 0 XXXXX X 0 0 1 XXXX XX 1 0 1 XXX XXX Bits3 4 0 1 X1 1 0 X2 Satate B BitsS function Bits0 gross 0 net 1 Bits1 Symbol positive 0 negative 1 Bits2 Overload or under zero 1 Bits3 dynamic 1 Bi...

Page 25: ...st Mode C18 3 In this mode the indicator can receive ASCII commands as listed below Command Name Function 1 Tare Save and clear tare 2 Zero Zero gross weight 3 Print Print the weight 4 G W N W Read gr...

Page 26: ...nding Mode C18 5 STX POL XXXXX XX LB KG GR NT STAT CR LF Polarity SP Positive Negative Weight Data Gross Net GR Gross NT Net Line Feed Start Transmission Units lb punt kg kilogram pcs pieces Status SP...

Page 27: ...the scale s internal 9VDC power supply will need to be on the same ground plane for the output current to be synced and measured correctly Please refere to the figure below as a visual clarification o...

Page 28: ...d HOLD key to go into configuration mode Go to C32 and press PRINT key The display should show out 4 and output should be at 4mA Press the up down arrow keys and the out will increase decrease Note X...

Page 29: ...trol function Output overload control signal Out2 Turn on compliance control function Output compliance control signal Out3 Turn on under load control function Output under load control signal Out4 Tu...

Page 30: ...See Q A section Err1 Weight is not used during calibration or the weight is above the maximum capacity Use correct weight within the defined range Err2 Weight is below the minimum requirement during c...

Page 31: ...to 100 if possible using the up arrow key 7 If 100 is not available change to 20 8 Press PRINT key to enter your selection 9 The screen will display C21 10 Press PRINT key to enter C21 parameter 11 Ch...

Page 32: ...You can send an email to sales usameasurements com for any questions about your sales or support Alternatively call 800 711 2237 for sales and support Visit our website at usameasurements com...

Reviews: