Please note:
baud rate is selected by
USER-COM1
/
2-BAUDRT
,
8N1
byte format is fixed,
Hex data (0x00 – 0xff)
are used.
7.8.2.
Press down
HOLD/SETUP
key until
CONFIG
is shown, using
keys to select
TEST-COM1.RD
or
TEST-COM2.RD
item, press
TARE/PRESET
to enter test COM1/2 receiving function, and
rd1
.--
or
rd2
.--
will be
displayed first.
7.8.3.
In this mode, received hex data (0x00 – 0xff) will be displayed on -- position.
7.8.4.
Press
HOLD/SETUP
key to return to last menu item, press
ZERO/ON/OFF
key to prepare to exit this mode
7.9.
Serial Port1/2(COM1/2) Transmitting Test
7.9.1.
Before test the transmitting function of
COM1
or
COM2
, a cable is need to connect a PC and this instrument ,
and a software be similar with Super Terminal of Windows is also need to run on PC to receive bytes from this
instrument. Please note:
baud rate is selected by
USER-COM1/2-BAUDRT
,
8N1 byte format is fixed
, Hex data
(0x00 – 0xff) are used.
7.9.2.
Press down
HOLD/SETUP
key until
CONFIG
is shown, using
keys to select
TEST-COM1.TD
or
TEST-COM2.TD
item, press
TARE/PRESET
to enter test COM1/2 transmitting function, and
td1
.--
or
td2
.--
will be
displayed first.
7.9.3.
In this mode, transmitted hex data (0x00 – 0xff) will be displayed on -- position, and
keys can
be used to modify transmitted data.
7.9.4.
Press
HOLD/SETUP
key to return to last menu item, press
ZERO/ON/OFF
key to prepare to exit this mode.