9. Troubleshooting

Symptoms

Troubleshooting method

Failure of Ultrasonic wave

① 

Check whether the power cable has been connected 
properly (by inspecting operation of power indicator lamp).

② 

Check whether PCB has been connected properly to the 
power socket.

③ 

Check whether the fuse tube is broken.

Weak ultrasonic wave

① 

Check whether the power voltage is normal.

② 

Inspect whether the energy converter get loose or there 
is any shock absorption material or work piece under the 
tank.

③ 

Check whether cleaning fluid temperature is normal 
(recommended operating temperature is higher than 30 
degree) or there are many air bubbles in the fluid or on the 
tank wall (repeated power-on/off will reduce the bubbles).

④ 

Check whether the solvent is added excessively.

No heating

① 

Check whether target temperature is too lower (lower than 
current actual temperature).

② 

Check whether the heater cable is disconnected or gets 
loose.

③ 

Check whether heater is broken (due to shortage of 
cleaning fluid).

Electric leakage

① 

Check whether the machine has absorbed damp or 
corroded.

② 

Check whether cable tray to each component is damaged, 
loosened or broken.

③ 

Check whether the earth wire is in good condition.

Summary of Contents for USA-UC-30L

Page 1: ...Version 2020 Ultrasonic Cleaner USA UC 30L...

Page 2: ...l Failure to follow the instruction in this manual may cause wrong operation 1 Principle of ultrasonic cleaning 3 2 Precautions 3 4 Specification 5 3 Product features 4 5 Scope of application 5 6 Inst...

Page 3: ...s to it This product contains dangerous voltage so do not start it at a high temperature or humid location to avoid electric shock This product must be operated after inject liquid otherwise it may be...

Page 4: ...shock or fire and keep it away from any location with high temperature or humidity For reason of personal safety do not dismantle the enclosure for repair or change on your own Proprietary MCU sweep u...

Page 5: ...eaning disinfection and sterilization of medical devices and lab glassware Semiconductor Fine cleaning of semiconductor chips Timepiece and jewelry Removal of grease dirt scale and polishing paste Che...

Page 6: ...ed on properties of work piece to be cleaned Properly and firmly connect the plug to a three pin power socket which is grounded reliably to ensure safety 7 Operation instructions Timing Min Power Swit...

Page 7: ...turns on press the button again to stop operation and operating indicator lamp goes out 8 Factors influencing performance Ultrasonic power density Higher power density results in better cleaning perf...

Page 8: ...whether cleaning fluid temperature is normal recommended operating temperature is higher than 30 degree or there are many air bubbles in the fluid or on the tank wall repeated power on off will reduce...

Page 9: ...tank for any leakage Check the joint between drain pipe and cleaning tank for leakage For looseness of ultrasound wave converter or burnout of power tube contact the manufacturer immediately The ultr...

Page 10: ...Memo...

Page 11: ...Memo...

Page 12: ...11730 Globe St Livonia MI 48150 Call us at 734 855 4890...

Reviews: