background image

- 7 -

Operation Instructions:

① 

Load cleaning fluid into the tank and connect the power supply (which shall be grounded 
reliably). After power-on, current actual temperature is displayed and the default 

temperature is 20℃. The ultrasound wave operating time is 99 minutes by default. 

② 

Adjustment of ultrasound operating duration: Press time setting button, one push to 
increase one minute; long press it to increase fast, in a loop of 1-99 minutes. 

③ 

Adjustment of operating temperature: Press temperature setting button, one push to 
increase one minute; long press it to increase fast, in a loop of 1-80 degree. 

④ 

After target temperature is set properly, press heating ON/OFF button to start heating tank 
(heating system is started and heating indicator lamp turns on when target temperature 
is higher than current actual temperature; heating system is stopped when target 
temperature is lower than current actual temperature), and press it again to stop heating.  

⑤ 

After operating duration is set properly, press timer ON/OFF button to start operation 
and the operating indicator lamp turns on; press the button again to stop operation and 
operating indicator lamp goes out. 

8. Factors influencing performance   

① 

Ultrasonic power density: Higher power density results in better cleaning performance and 
higher cleaning speed. Higher power density is appropriate for cleaning of work piece with 
stubborn dirt, and lower power density is suitable for any precision work piece (generally 
ultrasonic power density ranges between 0.01-0.02w/ml). 

② 

Ultrasonic Frequency: Lower frequency results in better cavitation performance. Higher 
frequency contributes to better refraction/ reflection performance. Lower frequency should 
be applied for cleaning of simple surface while higher frequency for cleaning of complex 
surface with deep or inaccessible hole. (80KHz, 120KHz)

③ 

Cleaning temperature: Ultrasound wave will present the best cavitation performance at a 

temperature range between 40℃~50℃. Higher cleaning temperature facilitates breakdown 
of  dirt,  while  a  temperature  above  70℃~80℃  may  impair  effect  of  ultrasound  wave  and 

deteriorate cleaning performance.

④ 

Cleaning duration: Longer cleaning duration presents better cleaning performance (except 
for special material).

⑤ 

Other influential factors include type and properties of cleaning fluid and the dirt to be 
removed.

Summary of Contents for USA-UC-30L

Page 1: ...Version 2020 Ultrasonic Cleaner USA UC 30L...

Page 2: ...l Failure to follow the instruction in this manual may cause wrong operation 1 Principle of ultrasonic cleaning 3 2 Precautions 3 4 Specification 5 3 Product features 4 5 Scope of application 5 6 Inst...

Page 3: ...s to it This product contains dangerous voltage so do not start it at a high temperature or humid location to avoid electric shock This product must be operated after inject liquid otherwise it may be...

Page 4: ...shock or fire and keep it away from any location with high temperature or humidity For reason of personal safety do not dismantle the enclosure for repair or change on your own Proprietary MCU sweep u...

Page 5: ...eaning disinfection and sterilization of medical devices and lab glassware Semiconductor Fine cleaning of semiconductor chips Timepiece and jewelry Removal of grease dirt scale and polishing paste Che...

Page 6: ...ed on properties of work piece to be cleaned Properly and firmly connect the plug to a three pin power socket which is grounded reliably to ensure safety 7 Operation instructions Timing Min Power Swit...

Page 7: ...turns on press the button again to stop operation and operating indicator lamp goes out 8 Factors influencing performance Ultrasonic power density Higher power density results in better cleaning perf...

Page 8: ...whether cleaning fluid temperature is normal recommended operating temperature is higher than 30 degree or there are many air bubbles in the fluid or on the tank wall repeated power on off will reduce...

Page 9: ...tank for any leakage Check the joint between drain pipe and cleaning tank for leakage For looseness of ultrasound wave converter or burnout of power tube contact the manufacturer immediately The ultr...

Page 10: ...Memo...

Page 11: ...Memo...

Page 12: ...11730 Globe St Livonia MI 48150 Call us at 734 855 4890...