![US endoscopy 710302 Operator'S Manual Download Page 31](http://html.mh-extra.com/html/us-endoscopy/710302/710302_operators-manual_1063369031.webp)
CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR Operator’s Manual
Page 30
Condition
Possible Causes
Field Checkout
My CO
2
EFFICIENT
ENDOSCOPIC
INSUFFLATOR does not
deliver CO
2
.
What could be the problem?
CO
2
EFFICIENT
ENDOSCOPIC
INSUFFLATOR flow rate
incorrect.
.
To check the CO
2
EFFICIENT ENDOSCOPIC
INSUFFLATOR
’s ability to deliver CO
2
to the patient,
follow these verification steps:
After powering-up and making sure that your
CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR is
connected to the CO
2
supply tank, depress the VOLUME
RESET button, zero will appear in the VOLUME LITERS
display.
Set the mode selector switch to FREE FLOW. Disconnect
the Tubing Set if connected. Press the FLOW STOP/RUN
button to start.
Using a wrist watch with a second hand or a stop watch,
allow the unit to deliver 3.0 liters of CO
2
. The unit should
reach this volume between 50 and 70 seconds. During this
time interval, place your finger near the Gas Output. You
should feel CO
2
exiting from the CO
2
EFFICIENT
ENDOSCOPIC INSUFFLATOR.