EFFICIENT ENDOSCOPIC INSUFFLATOR Operator’s Manual
Page 16
WARNING: Do not allow liquid CO
2
to enter the unit. This can be prevented by assuring that
the CO
2
supply tank is maintained in a vertical upright position at all times.
USE ONLY MEDICAL GRADE CO
2
SIZE “D” OR “E” supply tanks.
Before each use the following procedures or inspections should be performed:
Visually inspect the CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR for external
signs of damage.
Inspect the electrical connections. Do not use if inspection reveals damage.
Before connecting the power cord to the wall outlet, make sure the main power
switch is off and that the voltage is correct.
Inspect the connection to the CO
2
supply tank, to assure it is intact and tight.