CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR Operator’s Manual
Page 18
10.3 TUBING SET CONNECTION
Insert the connector on the Tubing Set to the Gas Output connection on the unit’s
front panel. It is important to use only US Endoscopy manufactured high flow tubing
labeled as US Endoscopy CO
2
Endoscopic Tubing Set (that includes a ≤0.1
micron
hydrophobic filter) designed to provide optimum performance. Always inspect every
Tubing Set to make sure there are no signs of damage. If such a condition exists, do
not use the Tubing Set. See Tubing Set instructions for use.
10.4 SETTING MODE OF OPERATION
To set the mode of operation, use the rocker switch on the front panel to select
between Managed Flow and Free Flow.
10.5 RESET CO
2
VOLUME
Assure that the Volume Liters Display reads zero prior to starting the procedure. To
reset the Volume Liters Display, press the Volume Reset button.
NOTE: Do not press Volume Reset button once procedure has started.
10.6 CONNECTION TO ENDOSCOPIC SYSTEM
The Tubing Set connects into the air/water line of the endoscopic system.
Depending on the Endoscope manufacturer there are different connections to the
air/water supply line. Each connection has a Luer adaptor and some examples are
provided below.
Figure 10.1 CO
2
Adaptor for Pentax Endoscope