18
Soak the handpiece with disinfectant water such as iodine, alcohol, glutaraldehyde, etc.
Put it into the oven or microwave oven to bake at high temperature.
4.2. Tip
The tips can be sterilized in high temperature and pressure environment.
4.3. Torque wrench and Endo wrench
1)
Torque wrench and Endo wrench can be sterilized in high temperature and pressure environment.
2)
It is strictly forbidden to sterilize in the following ways:
Put it in the solution and cook.
Soak it with disinfectant water such as iodine, alcohol, glutaraldehyde, etc.
Put it into the oven or microwave oven to bake at high temperature.
NOTICE:
We will not be responsible for the damage to the torque wrench and endo wrench directly or indirectly caused by the
improper use of the above methods.
4.4. Cleaning Tips, Torque wrench and Endo wrench
Tips, Torque wrench and Endo wrench can be cleaned by the ultrasonic cleaning machine.
5.
Troubleshooting
Error
Possible causes
Solutions
Turn on the power switch, there
is no indicator light on the panel
Poor contact of power plug
Plug in the power plug tightly
Internal fuse not working
Contact your local dealer or our company
There is no vibration or no water
spray at the tip after stepping on
foot switch
Poor contact of wired foot switch
Plug in the plug of foot switch tightly
Batteries are out of power of wireless foot
switch
Change new batteries
Foot switch not working
Match the foot switch again(please refer to paragraph 2.6)