14) After completing scaling, keep working for 30 seconds under the water supply to clean the handpiece and tip.
15)
Remove the tip and then pull out the handpiece for sterilization.
NOTICE:
1)
Please do not pull out the handpiece when stepping on the foot switch and the tip vibrates.
2)
If do not use the wireless foot switch for a long time, please remove the battery.
3)
Please do not move or turn over the device when it is working.
3.4. Endo function
1)
Fix the holder of endo file to the handpiece with endo wrench (See Figure 2.6 ).
2)
Unscrew the nut of the holder .
3)
Insert the endo file into the hole in front of the holder.
4)
Tighten the nut of the endo file with endo wrench.
5)
Press "E", the "E" indicator lights up.
6)
When the Endo function is selected, slowly extend the endo file into the patient's root canal, step on the foot
switch and start the root canal washing. Adjust the power of root canal washing as needed.
NOTICE
:
1)
Tighten the endo tip.
2)
Tighten the nut of the holder.
3)
Do not apply heavy pressure when root canal washing.
4)
Do not step on the foot switch when the endo file is not placed in the root canal.
5)
It is recommended that the power adjustment start from the 1st scale and slowly increase to the
3rd scale when using the Endo function.