VERSAMET 4 INVERTED METALLURGICAL MICROSCOPE
UNITRON
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com
16
OPERATION
(continued)
Changing from Brightfield
To Darkfield Observation
(Fig. 22)
The Versamet 4 allows users to easily change from
brightfield observation to darkfield observation by
sliding the brightfield (BF) & darkfield (DF) lever
①
.
For brightfield observation, push the lever all the way
in (BF).
For darkfield observation, open both the aperture
diaphragm and field diaphragm levers to fully open,
and pull the brightfield & darkfield lever all the way
out (DF).
Using the Simple Polarizer
(Fig. 23)
The simple polarizer includes the polarizer and
analyzer.
To use, remove the dust plug from the polarizer slot
①
and insert the polarizer slider
②
into the slot as
shown. The polarizer is moved into the light path
when you hear it the second click.
NOTE:
when using the polarizer, remove the color
filter slider.
There are two analyzers: one fixed
③
and the other
360° rotatable analyzer
④
.
To use one of the analyzers, remove the dust plug
from the analyzer slot
⑤
and insert an analyzer
slider into the slot as shown.
The polarizer and analyzer are orthogonal when the
360° rotating analyzer slider is zero adjusted or the
fixed analyzer slider is used.
With the 360° rotating analyzer slider, you can
change the orthogonal state of the polarized light by
turning the dial
⑥
.
①
Fig. 22
①
Fig. 23
②
⑥
③
⑤
④
Summary of Contents for Versamet 4