Calibration of U137, U235 & 237 Series Weight Indicators.

Date:141027   B01950   3(10)

Miscellaneous functions.

13

*=0

=1 Difference (flow) calculated at a signal on the CTS input, J1:9. See page 9 and B01521. No count mode.
=2 Difference (flow) calculated every measurement cycle. See page 9 and B01521.  No count mode.
=3 Difference (flow) calculated every 100 measurement cycle. See page 9 and B01521.  No count mode.
+4 Not used.
+8 Not used.

14

*0 Setpoints are compared with displayed weight. For fast setpoints, use short measurement time in Cs08.

+1 Setpoints are used for over load and compared with gross value. On time delay is between 1 and 2 times

the measurement time according to Cs08. Off time delay is 8 times the measurement time.

+2 Not used.
+4 Not used.
+8 Not used.

15

*0 No battery timer.

+1 Leading zero blanking. For U235 series.
+2 Not used.
+4 Battery timer   64 measurement times time in Cs08.
+8 Battery timer  256 measurement times time in Cs08.

+12 Battery timer 1024 measurement times time in Cs08.

The timer is reset at button push or weight change more than 8 times the motion band.  U235 & U2372.

16

*=0 .

+1 Not used.
+2 Not used.
+4 Not used.
+8 Not used.

Calibration.

17

=0 x x x x x . 
=1 x x x x . x

*=2 x x x . x x

=3 x x . x x x
=4 x . x x x x  (Not display for U237)
=5 No decimal point.
+8 TARE is combined ZERO and TARE button. For U235 & U2373. All other buttons works as normal.

1

18

*=0 1  Verification scale interval (increment) e or e  at multiple range.

=1 2
=2 5
=3 10
=4 20
=5 50
+8 Only LED indicator is sent out on synchronous output. (G ives more stable display.)

19

*=0 Single range. Below the range change in % of max gross weight as function of min interval.

=1 2  ranges:     1  at  50% .

2  at  40% .

5  at  50% .

=2 3  ranges:     1  at  20  &  40% .

2  at  20  &  50% .

5  at  25  &  50% .

=3 4  ranges:     1  at  10  &  20  &  50% .

2  at  10  &  25  &  50% .

5  at  10  &  20  &  40% .

=4 5  ranges:     1  at  5  &  10  &  25  &  50% .

2  at  4  &  10  &  20  &  40% .

5  at  5  &  10  &  20  &  50% .

=5 6  ranges:     1  at  2  &  4  &  10  &  20  &  40% .

2  at  2  &  5  &  10  &  20  &  50% .

5  at  2.5  &  5  &  10  &  25  &  50% .

=6 Not used.
=7 Calibrated value with one additional digit and always 1 in interval. Not U1370 and U235.
+8 Multiple range change to range 1 is made only at zero indication. Range up is locked at stable weight.

Range indication. U137, U235: 1 no; 2 lower; 3 upper; 4 lower + upper right indicator on. U237: Range 1
none; 2 low; 3 +middle; 4 +high; 5 +low left; 6 +low right segment on, last digit. +, add to previous range.

1

Zero range is up to ±2%  of Max .

The advantage of more ranges is a faster scale and less sensitivity to tilt.
For legal scales the following data must be met in the test certificate of the load cell (See B01990):

n

1

max

Max /e  

#

 2.5Z.

     Z = E

/2DR. DR = Min dead load output return. If there is no Z in the load cell test

certificate, Z = 1000x may be used, where x is from load cell accuracy class designation Cx.

n

Max  may be set according to Cs21.

max

1

max

max

min

min

E

A

/

N/e  

#

 Y.

     E

 = Max load cell capacity. N = Number of load cells. Y = E

/v

. v

 = M in verifica-

min

tion scale interval. Y or v

 from the load cell test certificate.

max

exc

1

exc

1000

@

C/E

@

U

/N

@

e  

$

 0.6 

:

V.

     C = Output in mV/V. U

 = 10V except U2379, which has 5V.

20

*=0 Normal calibration for scales. Lower respective upper horizontal segments in the display are on when inside

AD-converter range below respective above the calibrated range.

=1 No zero functions. Full AD-converter range is used.
=2 No limits on zero range. Multiple range also works at decreasing weight.

Note! Can not be combined with Peak value mode Cs2: +4.

=3 Inverted calibration (for displacement level meters). No zero functions. Full AD-converter range is used.
=4 As Cs19:+8 but change to low range is made at zero or negative gross weight.
+8 If the last decimal at multiple range is a non-significant zero, it is blanked.

Summary of Contents for U137 Series

Page 1: ...21mV V W6 cut off There is a 3 pole filter with low ringing The response time is 1 6 seconds or 0 5 seconds with S2 1 2 and 3 on Keyboard functions at calibration The software calibration is performed...

Page 2: ...RS422 or in Cs1 3 4 RS422 is addressed by ASCII 124 1 to 14 CR No one is addressed by and character other than 1 to 9 CR is recommended 15 Always addressed Used for single RS4225 unit without address...

Page 3: ...3 4 ranges 1 at 10 20 50 2 at 10 25 50 5 at 10 20 40 4 5 ranges 1 at 5 10 25 50 2 at 4 10 20 40 5 at 5 10 20 50 5 6 ranges 1 at 2 4 10 20 40 2 at 2 5 10 20 50 5 at 2 5 5 10 25 50 6 Not used 7 Calibrat...

Page 4: ...change in AD value 25 Third calibration point 00000 Display data 3 Must be even number 000000 Internal AD value Cs25 is not used when this value is lower than that in Cs24 Full AD converter range is...

Page 5: ...both tilts and used in the formula The calibration can not be left with S1 when displaying angle These errors are caused by different angle between tilt sensor and load cell s and by the side force se...

Page 6: ...mm month and dd day Option 190 Time as hh mm ss where hh are hours mm minutes and ss seconds Option 191 Date as 20yy mm dd where yy are the two last digits of the year mm month and dd day Option 192 W...

Page 7: ...tation voltage 10V 1 However U2379 5V 1 Connect GND to metal hood of D sub J3 U237 series Power interface connector 9p D sub male B01910 1 2 3 4 5 6 7 8 9 0V Return for J3 2 Power V 0VD VB V after fus...

Page 8: ...25mA SYNC CK SYNC D1 SYNC D2 TARE PT RET PRINT TEST DTR Connect to metal hood of D sub RS232 B01910 U137 has inverted TTL output Use J3 for RS232 RS232 U137 has inverted TTL output Use J3 for RS232 R...

Page 9: ...be made in separate grounded metal boxes Beware of high temperature sources as the signal is very sensitive to temperature gradients The indicator must be well protected against high humidity dust and...

Page 10: ...ht Indicators Date 141027 B01950 10 10 Unisystem AB Bergebyv 24 Tel 0560 140 55 www unisystem se 685 34 TORSBY S te Fax 0560 101 25 info unisystem se Old display U2370 U2371 U2372 U2373 U2375 U2379 U2...

Reviews: