Event Output at the End of Graph Plotting
Each time a graph is plotted on screen, a pulse signal is output (the pulse width is 200msec). The
waveform can automatically be taken in through the RS-232C interface by connecting to the event
input of the DS400.
5-21.
RUN Output
External output “RUN” stops when abnormal state of indicated value is detected.
The conditions of stopping the output caused by abnormal state are as follows.
・
Sensor
±
error
・±
Over (Overflow error)
・
Overload (Overload error)
・
Zero error (Zero calibration error)
・
Span error (Span calibration error)
・
Zero limit (Digital zero limit error)
RUN out specifications are as follows.
RUN output spec
Normally, the signal repeating ON-OFF operation is output.
CAUTION
By the selection of each LOCK function, measurement work that can be
specified externally is limited to WORK00 to WORK07.
SET.
Operation
Page 2
→
→
System Setting
→
Event output
OFF
ON
200msec
The end of graph plotting
*If the time to start the next graph plotting is less than 200msec, the pulse is
forcedly turned OFF when the plotting starts.
Refer to "Error Display List"
on page 85 about the contents
of errors.
RUN
OFF
ON
250msec pulse
This signal turns off when an error occurs.
duty50%
RUN
OFF
ON
3.0sec
Error factor removal
By the error factor removal, ON-OFF
operation restarts in 3.0sec.
Summary of Contents for F377A
Page 1: ...18JUL2012REV 3 06 DIGITAL INDICATOR F377A OPERATION MANUAL ...
Page 5: ...What is RoHS 4 M E M O ...
Page 12: ...11 11 STANDARD ...
Page 13: ...12 12 M E M O ...
Page 94: ...93 93 EXPANSION ...
Page 95: ...94 94 M E M O ...
Page 97: ...Expansion Function 96 M E M O ...
Page 152: ......