background image

                     Service manual Phyaction V

 

 

 

 

 

Service manual version 0.1

 

 

 

Page 5

 

4.

 

Technical data  

 
This is a summary of the technical data as described in the user manual. 

4.1

 

General technical data 

 

  Dimensions Phyaciton V (w x h x d) 

267 x 95 x 270 mm 

  Weight Phyaciton V  

 

 

3,5 kg 

  Weight including accessories  

 

4,6 kg 

  Mains voltage    

 

 

100 - 240 VAC, 50 - 60 Hz 

  Maximum power consumption    

30 VA 

  Safety class  

 

 

 

Class I (earthed socket required) 

  Insulation  

 

 

 

Type BF (floating patient circuit) 

 

Fuses 

 

    2 

T2AL250V 

  Volume water reservoir   

 



180 ml 

  Working pressure continuous vacuum   38 - 320 hPa 

  Working pressure pulsation vacuum  

58 - 480 hPa 

 

Vacuum 

rhythm 

    1,5/1,5 - 1,5/4,5 s (on/off time)   

4.2

 

Environmental conditions 

 

  Temperature: 

 

 

 

+10 °C to +40 °C 

  Relative humidity  

 

 

30% to 75% 

  Atmospheric pressure    

 

700 hPa to 1060 hPa 

4.3

 

Transport and storage 

 

  Transport weight  

 

 

6,5 kg 

  Storage temperature  

 

 

-20 °C to +60 °C 

  Relative humidity  

 

 

10% to 100%, including condensation 

  Atmospheric pressure    

 

200 hPa to 1060 hPa 

  Transport classification   

 

Single piece by mail 

  

Summary of Contents for Phyaction V

Page 1: ...pply 5 2 6 Patient Current Routing 5 3 6 Vacuum 5 3 1 6 Control Loop 5 3 2 6 Pulsed Vacuum 5 3 3 7 End of Treatment 5 3 4 7 Water management 6 8 Safety inspection list yearly check up 7 9 Hardware lab...

Page 2: ...r of the Uniphy family it is built according to the requirements of the MDD security standards All functions of the unit are controlled by a microprocessor ensuring a high degree of reliability and sa...

Page 3: ...e able to provide the following data for each unit Instrument data Part and serial number Gymna has the original configuration of each unit The Phyaction V has following critical parts Ref 324597 Main...

Page 4: ...er name Customer address Instrument name Serial number Date of delivery Service activities Date Error complaint Service action Critical spare parts exchange list Critical part name Old serial number N...

Page 5: ...socket required Insulation Type BF floating patient circuit Fuses 2 x T2AL250V Volume water reservoir 180 ml Working pressure continuous vacuum 38 320 hPa Working pressure pulsation vacuum 58 480 hPa...

Page 6: ...el B by pressing a button that toggles the choice between traditional and vacuum electrodes 5 3 Vacuum When the vacuum control is turned up air is pumped out to create the vacuum until the set level i...

Page 7: ...avoidably droplets are sucked into the system These droplets do not only contain water but also fibers from the sponges flakes of skin and other substances present under the cups To prevent these to r...

Page 8: ...the keys and the display are undamaged Mains entry and power cord are undamaged The output connectors are undamaged The electrode connectors and cables are undamaged All the operation functions work T...

Page 9: ...ut The last 3 alphanumerical numbers will begin with A A A to Z Increase in case of a principle technology change Components are not anymore compatible with the previous layout All previous changes ma...

Page 10: ...ing static sensitive devices such as PC boards of the Phyaciton V the following precautions should be observed Persons should be earthed by means of a wrist strap Ground all electrical equipment work...

Page 11: ...pply PCB Unscrew the 4 screws of the power supply PCB and replace with a new PCB 8 3 Replacement of the Main PCB The main PCB is one board in production that is divided over the front of the unit To r...

Page 12: ...m the patient We have noticed that is some cases the vacuum is switched off even when there is current detected In order to solve this the unit needs to be modified Step 1 open the unit Step 2 remove...

Page 13: ...324597 MAIN PCB PHYACTION V VACO 200 324650 CABLE ASSY ET IN VACO500 PHYACTION V 324727 CABLE ASSY MAINS INLET SMPS 200 SERIES 324793 BOTTOM BOX PHYACTION V 324804 METAL TOP COVER PHYACTION V 324837 C...

Page 14: ......

Page 15: ......

Page 16: ......

Page 17: ...R247 ELVA003 LO200 V21 pcb Tue Dec 13 10 38 58 2005...

Page 18: ......

Reviews: