9.

 

Upgrades 

 

9.1

 

Vacuum pump switches of after 2 minutes even when current is flowing 

 
The Phyaction V has an automatic current sense circuit. If the unit doesn’t detect any current for 
more than 120 sec, the vacuum is switched off and the vacuum cups will be released from the 
patient. 
We have noticed that is some cases, the vacuum is switched off, even when there is current 
detected. In order to solve this, the unit needs to be modified. 

 

Step 1: open the unit 
Step 2: remove the main PCB from the unit 
Step 3: remove R247 from the main PCB (photo below). Be careful about the 2 adjacent 
capacitors. 

 

Remove R247 

 

 
Step 4: install a 1Mohm resistor between C233p+ and CON216p7 (photo below). 
 

 

Install 1Mohm resistor 

Step 5: change the hardware code into 1AAB 
Step 6: close the unit and perform a functional and safety test 

 

 

Summary of Contents for Phyaction V

Page 1: ...pply 5 2 6 Patient Current Routing 5 3 6 Vacuum 5 3 1 6 Control Loop 5 3 2 6 Pulsed Vacuum 5 3 3 7 End of Treatment 5 3 4 7 Water management 6 8 Safety inspection list yearly check up 7 9 Hardware lab...

Page 2: ...r of the Uniphy family it is built according to the requirements of the MDD security standards All functions of the unit are controlled by a microprocessor ensuring a high degree of reliability and sa...

Page 3: ...e able to provide the following data for each unit Instrument data Part and serial number Gymna has the original configuration of each unit The Phyaction V has following critical parts Ref 324597 Main...

Page 4: ...er name Customer address Instrument name Serial number Date of delivery Service activities Date Error complaint Service action Critical spare parts exchange list Critical part name Old serial number N...

Page 5: ...socket required Insulation Type BF floating patient circuit Fuses 2 x T2AL250V Volume water reservoir 180 ml Working pressure continuous vacuum 38 320 hPa Working pressure pulsation vacuum 58 480 hPa...

Page 6: ...el B by pressing a button that toggles the choice between traditional and vacuum electrodes 5 3 Vacuum When the vacuum control is turned up air is pumped out to create the vacuum until the set level i...

Page 7: ...avoidably droplets are sucked into the system These droplets do not only contain water but also fibers from the sponges flakes of skin and other substances present under the cups To prevent these to r...

Page 8: ...the keys and the display are undamaged Mains entry and power cord are undamaged The output connectors are undamaged The electrode connectors and cables are undamaged All the operation functions work T...

Page 9: ...ut The last 3 alphanumerical numbers will begin with A A A to Z Increase in case of a principle technology change Components are not anymore compatible with the previous layout All previous changes ma...

Page 10: ...ing static sensitive devices such as PC boards of the Phyaciton V the following precautions should be observed Persons should be earthed by means of a wrist strap Ground all electrical equipment work...

Page 11: ...pply PCB Unscrew the 4 screws of the power supply PCB and replace with a new PCB 8 3 Replacement of the Main PCB The main PCB is one board in production that is divided over the front of the unit To r...

Page 12: ...m the patient We have noticed that is some cases the vacuum is switched off even when there is current detected In order to solve this the unit needs to be modified Step 1 open the unit Step 2 remove...

Page 13: ...324597 MAIN PCB PHYACTION V VACO 200 324650 CABLE ASSY ET IN VACO500 PHYACTION V 324727 CABLE ASSY MAINS INLET SMPS 200 SERIES 324793 BOTTOM BOX PHYACTION V 324804 METAL TOP COVER PHYACTION V 324837 C...

Page 14: ......

Page 15: ......

Page 16: ......

Page 17: ...R247 ELVA003 LO200 V21 pcb Tue Dec 13 10 38 58 2005...

Page 18: ......

Reviews: