4.1.2 Set Serial Number: 
 
 
 

 
 
 
 

1. 

In this screen the display will show the current serial number of the instrument.

  

2.   The cursor will be blinking at the first digit of t

he “Serial Number”, make desire changes 

and press ENT key to confirm.

  

 

 

4.1.3 Date Time Setting: 

 
 

 
 
 
 
 
 

        

In a similar way, one can change the Date & Time. 

 

Note:

 If the displayed date & time or any of the above explained parameters are not to 

be changed, one can press the 

SET Key ● to escape without altering the current values. 

 

 

4.2  CALIBRATION MODE: 

 

In  the  menu  screen,  press  the  DOWN

 

Key 

,  the  instrument  moves  to  the  calibration 

mode.    After  Successful  authentication  (Password 

–  321),  the  device  is  ready  for 

calibration.  

   

 

   Calibration of the instrument involves two steps: 
 
1) 

Zeroing the instrument in clean air. 

 

 

∆ 

IMPORTANT

The  instrument  should  be  zeroed  in  clean  air.  If  you  are  not  sure 

about  the  air  quality,  it  is  recommended  to  use  the  air  from  compressed  clean  air 
cylinders available for the purpose. 

 

 

2)  Applying gas of known concentration, so that the reading matches with the calibration 

gas concentration. 

 

 

4.2.1 Calibration Process:  

 

1.  In the calibration mode, the first page is the Zero Calibration, 

“Apply Zero Gas”

 Prompt 

will be displayed and instrument will take samples from air for 30 seconds, continuously 
showing counts and remaining sampling time on display. 

 

2.  Instrument  will  be  zeroed  only  if  count  value  is  in  between

  50  to  999.

  After  zeroing 

successfully  ZERO  CAL  DONE  prompt  will  appear  otherwise  ZERO  CAL  FAIL  prompt 
will be displayed. 

Set Date: 
DD/MM/YY 
Set Time: 
HH:MM:SS 

 

Serial Number 
A – XXXX (PM) 

Summary of Contents for FumiSense Pro UNIPHOS-250

Page 1: ...S 250 PM FumiSense Pro INSTRUCTION MANUAL Uniphos Envirotronic Pvt Ltd Readymoney Terrace 167 Dr Annie Besant Road Worli Mumbai 400018 India Tel 91 22 2493 0681 Fax 2493 8826 Email gasdetection unipho...

Page 2: ...ment and the full details of operation for measurement calibration parameter setting etc To ensure a long trouble free performance please take due care and ensure proper maintenance of your monitor as...

Page 3: ...G TYPE SAMPLING TIME 10 4 1 2 SET SERIAL NUMBER 11 4 1 3 DATE TIME SETTING 11 4 2 CALIBRATION MODE 11 4 2 1 CALIBRATION PROCESS 11 4 3 SAMPLING MODE 12 4 4 DATA DOWNLAODING MODE 14 4 4 1 TO DOWNLOAD R...

Page 4: ...demands monitoring of the fumigation process during the entire period of fumigation For this application UNIPHOS 250 PM FumiSense Pro is suitable This manual gives the full description of the UNIPHOS...

Page 5: ...pack Being a microprocessor based unit it has data logging facility and can store the data with a real time clock into the memory There is also a provision to download the stored data on the computer...

Page 6: ...strument s memory CAUTION One should be careful before pressing DOWN key in the Data Downloading Mode as the data logged so far into the internal memory of the instrument will be lost permanently 5 SE...

Page 7: ...for easy downloading of the data stored in the data logger The other end is to be connected to the COM port of the PC The settings required for data downloading are described in point 4 4 Data Downloa...

Page 8: ...rovided for charging the instrument Plug the charger to 110V 230VAC 50 60Hz power outlet and the other end to the instrument and turn it ON Charge the battery for 3 4 hours The Battery charge status c...

Page 9: ...g it for measurement 3 1 INITIALIZATION Once instrument is turned ON it goes into the initialization and following prompts will appear in the sequence on the display Instrument Turn ON Instrument Mode...

Page 10: ...be changed move the cursor under the digit and press the UP Key Using the UP Key the digits can be changed from 0 to 9 3 Press the ENTER Key to confirm the changed digit and to switch over to the nex...

Page 11: ...the instrument involves two steps 1 Zeroing the instrument in clean air IMPORTANT The instrument should be zeroed in clean air If you are not sure about the air quality it is recommended to use the a...

Page 12: ...k for SPAN GAS VALUE 6 Instrument will be calibrated only if counts value for applied span gas is in valid range i e From 2000 to 4030 display counts ideally 3000 7 After Span Calibration process SPAN...

Page 13: ...key to enter into the sampling mode 5 In continuous sampling mode instrument will start taking sample until the SET key is pressed to stop the sampling Note In the continuous sampling mode data can be...

Page 14: ...sensor drift and instrument needs to be calibrated by following procedure described in 4 2 4 4 DATA DOWNLOADING MODE 1 In the menu page when the ENTER key is pressed instrument will enter into the dat...

Page 15: ...al communication that is included in Microsoft Windows 98 and higher versions It can be configured as follows 1 Click the START button on the PC 2 Go to Programs Accessories Communications HyperTermin...

Page 16: ...main menu page 5 How to Monitor Battery Charge Level The battery charge level can be viewed on the LCD OLED panel when the instrument is in main Data Downloading page The LCD OLED will indicate the ba...

Page 17: ...4 Alphanumeric LCD OLED Display 3 Inbuilt pump Flow rate 500 1000 ml min Power 4 5 V DC 35 mAH 4 Power supply Inbuilt 7 4 V Li Ion rechargeable battery 5 Weight 0 57 Kg 1 256 lb 6 Construction materi...

Page 18: ...but also depends on the accuracy of the calibrating gas 8 Drift Less than 2 of signal per month at constant temperature and pressure 9 Operating temperature 0 to 40 oC 10 Operating pressure Ambient 1...

Page 19: ...in a target gas free clean air 5 NEVER expose the sensor to gas concentration higher than its prescribed range 6 DO NOT expose the monitor to excessive positive or negative pressure It affects the acc...

Page 20: ...ng Problem Possible cause Solution to the problem The instrument does not turn ON Low Battery Recharge battery Pump making abnormal noise There may be blockage in the gas sampling tube or in the nozzl...

Page 21: ...0 Calibration Schedule Instrument No Calibration Gas Sr No Date Standard Gas Conc Instrument reading Remarks Next Calibration Due Date Note Calibration of the instrument is recommended on quarterly ba...

Page 22: ...22 11 CONCLUSION REMARKS...

Page 23: ...usly compromise the user s life The device must be used only in compliance with the applicable regulations otherwise the safe operation of the instrument is not guaranteed After completion of life cyc...

Page 24: ...to the Seller s factory freight prepaid and will be returned at the Seller s risk of loss to the buyer s premises freight prepaid The foregoing is IN LIEU OF all other warranties and is subject to th...

Page 25: ...aon Dist Valsad 396 108 India Tel 91 9909994042 91 7574839945 Website www uniphos she com E Mail info uniphos envirotronic com After Sales support service uniphos envirotronic com Issue 0 Rev 0 Date 0...

Reviews: