Operation
59
Main menu - Options - Calibration - Automatic calibration
Day is a list field (Son, Mon, Tue, ...) while time and cycle are value fields.
The device itself determines the duration of purge time for the calibration gas. For
this purpose it uses a stability criterion. This criterion is reached after 6 to 10 min,
depending on the gas. Afterwards the device automatically switches back to
process gas. If the criterion is not reached, the calibration is aborted after 10 min.
The abortion is documented in the events list.
Attention: After a device restart or a change of calibration gas, the calibration might
need to be repeated 2-3 times. Successful calibration is only possible after the
pulse line to the calibration gas has been purged sufficiently and the entire air has
been displaced out of the line.
Summary of Contents for CWD2000 EX
Page 1: ...April 2015 V1 04 Original operating instructions Combustion calorimeter CWD2000 EX ...
Page 6: ...Table of contents 6 ...
Page 9: ...Technical data 9 ...
Page 10: ......
Page 12: ...Table of contents 12 ...
Page 20: ...Table of contents 20 2 3 ...
Page 21: ...Safety devices 21 4 5 ...
Page 38: ...Table of contents 38 ...
Page 42: ...Table of contents 42 ...
Page 48: ...Table of contents 48 9 6 Menu structure ...
Page 81: ...Operation 81 ...
Page 82: ......
Page 86: ...Table of contents 86 ...
Page 90: ...Table of contents 90 ...
Page 98: ...Table of contents 98 ...
Page 100: ...Table of contents 100 ...