Operation
55
Main menu - Options - I/O - Digital inputs
Signal and zero position are list fields. These inputs can start a calibration or switch
a holding amplifier on or off. The inputs are queried once per second. i.e. a change
must be pending for at least one second in order to be detected by the system.
Signals
Zero position
Contact
Combustion calorimeter
Start calibration
high
open
Calibration starts
closed
No calibration
low
open
No calibration
closed
Calibration starts
Start measurement
high
open
Measurement starts
closed
Measurement stops
low
open
Measurement stops
closed
Measurement starts
Signals hold
high
open
mA hold
closed
mA online
low
open
mA online
closed
mA hold
Calibration cancel
high
open
Calibration cancellation
closed
No calibration cancellation
low
open
No calibration cancellation
closed
Calibration cancellation
Summary of Contents for CWD2000 EX
Page 1: ...April 2015 V1 04 Original operating instructions Combustion calorimeter CWD2000 EX ...
Page 6: ...Table of contents 6 ...
Page 9: ...Technical data 9 ...
Page 10: ......
Page 12: ...Table of contents 12 ...
Page 20: ...Table of contents 20 2 3 ...
Page 21: ...Safety devices 21 4 5 ...
Page 38: ...Table of contents 38 ...
Page 42: ...Table of contents 42 ...
Page 48: ...Table of contents 48 9 6 Menu structure ...
Page 81: ...Operation 81 ...
Page 82: ......
Page 86: ...Table of contents 86 ...
Page 90: ...Table of contents 90 ...
Page 98: ...Table of contents 98 ...
Page 100: ...Table of contents 100 ...