Appendix A - OPERATING ERRORS
When the PROLOGGER detects certain error conditions it sets flags in
Register Address 3.
Flag
Meaning
Result
CS
Checksum error or other inconsistency found in the
PROLOGGERs Configuration Table or firmware
Fatal
BT
Battery is too low for safe operation.
Fatal
OS
Operator Stop. logger entered Sleep mode by
command (not due to detected error.)
Fatal
MB
Lithium memory backup battery needs replacement or is
not installed (default).
Warning
B1 B0
00 Battery OK, 01 Battery Low,
10 Battery Flat, 11 Battery Dead
Warning
RT
User Log Program exceeded maximum runtime and was
prematurely terminated (no EXIT instruction or
corrupted/incorrect Log Program.
Warning
SC
Scan counter fault detected.
Warning
Error flags are divided into WARNING and FATAL errors. FATAL errors
cause the PROLOGGER to stop operating and revert to sleep mode.
WARNING errors are for information only.
To wake a logger from sleep mode the logger must be connected to a
computer and the RTS signal presented. This powers on the logger but does
not allow signal scanning or log programs to run, however normal RS-232
communications can occur after the prompt (*CR).
A logger RESET will also set the logger from sleep to operating state.
PROLOGGER Hardware
1
5
3
2
4
1
RT
0
SC
B1
MB OS
B0
BT
6
CS
7
BITS
Register Address 3
(ERROR FLAGS)