Service Manual UTS1000/3000 Series
Instruments.uni-trend.com
13
/
25
5.
Disassembling
Removal tool
Use the following tools to remove or replace modules in the spectrum analyzer.
Item Number
Tool name
Description
1
Torque screwdriver
Model see disassembly steps
2
Torque wrench
Used to disconnect RF connectors or ports
3
Upholstered
Prevents damage to the screen and knobs when removing the
grid
4
Anti-static
environment
To prevent damage to devices caused by static electricity,
wear properly grounded anti-static clothing, wrist straps and
foot straps; effective anti-static mats
Remove the dust cover
The following procedure describes the removal and replacement of the rear cover.
prerequisites:
•
To prevent electrostatic damage to components, wear a properly grounded antistatic wrist and foot
strap during installation, and use an antistatic mat in a tested antistatic environment.
Steps:
1.
Use a T8 Torque screwdriver to remove the 4 screws that fix the dust cover and the rear cover of the rear
cover assembly, as shown in the figure below:
2.
Remove the dust-proof net, as shown in the figure below, and then clean the dust-proof net.