Service Manual UTS1000/3000 Series
Instruments.uni-trend.com
6
/
25
Working
environment
This device is intended for indoor use in a clean and dry environment with an ambient
temperature range of
0
℃
~
+40
℃
。
Do not operate the device in explosive, dusty or humid atmospheres.
Do not operate in
a wet
environment
Avoid the risk of short circuit or electric shock inside the instrument, and do not
operate the instrument in a humid environment.
Do not operate in
a flammable and
explosive
environment
To avoid instrument damage or personal injury, please do not operate the instrument
in a flammable and explosive environment.
Caution
Abnormal
situation
If you suspect that the product is malfunctioning, please contact the maintenance
personnel authorized by Unilever for testing;
The relevant person in charge of Unitech must carry out any maintenance,
adjustment, or parts replacement.
Cooling
requirements
Do not block the ventilation holes located on the sides and rear of the device;
Do not allow any foreign objects to enter the device through ventilation holes, etc.;
Ensure adequate ventilation, leaving at least 15 cm of clearance on the sides, front,
and rear of the unit.
Pay attention to
handling safety
In order to prevent the instrument from slipping during transportation and causing
damage to the buttons, knobs, or interfaces on the panel of the instrument, please pay
attention to the safety of transportation.
Maintain proper
ventilation
Poor ventilation can cause the temperature of the instrument to rise, which can cause
damage to the instrument.
Keep well-ventilated when in use, and check the vents and fans regularly.
Please keep it
clean and dry
To avoid dust or moisture in the air from affecting the performance of the instrument,
please keep the surface of the product clean and dry.
Notice
Calibration
The recommended calibration cycle is one year. Calibration should only be
performed by suitably qualified personnel.
Environmental requirements
This instrument is suitable for the following environments:
●
Indoor use
●
Pollution Degree 2
●
When operating: the altitude is lower than 3000 meters; when not operating: the altitude is lower than
15000 meters price
●
Under the premise of no special instructions, the operating temperature is 0 to
﹢
40
℃
; the storage
temperature is -20 to
﹢
70
℃
●
Humidity operates as Below +35
℃
≤90% relative humidity, non-operating humidity is +35
℃
~
+40
℃
≤60% relative humidity
There are vents on the rear panel and side panels of the instrument, please keep the air circulation through
the vents of the instrument case. Do not place the analyzer side-by-side with any other instrument that
requires side-by-side ventilation. Make sure that the exhaust port of the first instrument is away from the air