69
UTD4000 Four-channel User Manual
Unlock
Locked
k
Notes :
Auto Calibration :
UTILIITY
: In Auto mode the setup will go back to
default for auto setup.
: In Auto mode the current status is locked for
auto setup.
For unloc or locked status see the table below 2-30.
The adjustment basis for the trigger signal source in
Auto mode :
1. If there is input signal in only one channel, that
channel will be the trigger source.
2. When more than one channels have input signal,
the system will search for trigger signal from CH1.
The first channel where signal is detected will be
the trigger source.
3. If no signal is detected in any channel, CH1 will be
the trigger source
You can correct measurement
errors caused by environmental changes with the
auto calibration function. This process can be run
selectively when necessary. To make calibration more
accurate, power on your oscilloscope and allow 20
minutes to warm up. Then press the [
] button
(help function) and follow on-screen instructions.
.
Important Points :