background image

UTD1000L User Manual

65

Picture 6-21 Time Parameter Measurement

The time parameters that oscilloscope can

measure automatically include

Rise time: duration for waveform rising from 10%

to 90%

Fall time: duration for waveformfalling from 90%

to10%

Positive pulse width: pulse width when positive

pulse is at 50% amplitude

Negative pulse width: pulse width when negative

pulse is at 50% amplitude

Rise delay: delay time of A to B rise edge

Fall delay

delay time of A to B fall edge

Positive duty ratio: ratio of positive pulse width with

period

Negative duty ratio ratio of negative pulse width

with period

Press key RECORD to open waveform record

function menu, two independent menus as shown

in the following picture

:

:

:

Storage and Record

Record Function

Summary of Contents for UTD1025CL

Page 1: ...ent correctly please read the manual thoroughly and especially the Safety Notes part before using the device If you have read through the manul you are recommended to keep the menu properly with the instrument together or at the place you can read anytime in order to read it in the process of future use ...

Page 2: ... How to Conduct General Check Check If Any Damage Resulted By Transportation Check Accessory Check the Complete Apparatus Oscilloscope Connection Compensation Signal Output Connection of the Apparatus Oscilloscope panels and keys Chapter IV Primary Application of Oscilloscope Connect Appatatus to the Electricity Power nterface Instruction I Page General safety Overview 4 5 Preface 1 5 6 6 6 6 7 7 ...

Page 3: ...rizontal System Automatic Setting of Waveform Display Stop Restore Data Collection Trigger Frequency Count Setting Screen Brightness Adjustment Store Bitmap Online Help Set Single Trigger Check Oscilloscope Status Page 14 14 15 15 16 17 18 18 19 21 21 22 23 23 23 24 Operation Methods for Function Menu Reset Oscilloscope UTD1000L User Manual Menu Hiding ...

Page 4: ...re Diode Continuity Test Measure Capacitance Display Spike Pulse by the Peak Value Test Function Use Persistence to Show Waveform Select AC Coupling Reverse Polarity of the Displayed Waveform Automatic Setting for Small Signal with DC Bias Connect Apparatus Page 25 Chapter V Use Multimeter UTD1000L User Manual Check Oscilloscope System Information 25 26 27 28 28 29 29 30 32 32 32 33 33 34 34 35 ...

Page 5: ...tails about Using Oscilloscope Horizontal Sysmem Setting AC Vertical System Setting Trigger System Setting Collection Mode Setting Display Set Parameter Measurement Setting Storage and Record Record Function Storage Function Page 65 68 36 36 37 39 41 41 42 43 43 46 49 57 58 60 65 Measure AC Voltage Measure DC Voltage UTD1000L User Manual ...

Page 6: ...oduct Program Upgradation Program Upgradation Conditions Program Upgradation Appendix A Performance Index Appendix B Maintenance and Cleaning 69 71 I Preparations before Upgradation Page Chapter VII Troubleshooting Chater VIII Service and Support Chapter IX Appendix UTD1000L User Manual 73 74 75 78 79 81 81 81 81 81 84 84 92 ...

Page 7: ...er life noting statement indicates the conditions and activities that might cause damage to the product and other property The following terms might be on the product it means reading the symbol might be harmful to you immediately it means reading the symbol it means damage might be caused to the product or other property The following symbols might be on the product 1 Please pay attention to the ...

Page 8: ...lloscope or use the accessories designated by UNI T applicable to oscilloscope apparatus series product Before using check if any mechanic damage to the multimeter test pen oscilloscope probe and accessories and please replace it if any damage were found To remove all unused test pen probe an accessories To insert adapter into the AC socket and connect it with the oscilloscope In measurement in th...

Page 9: ...e the oscilloscope with the designated method The voltage rating value mentioned in the Warning information is the limit value of adapter Working Voltage they represent the Vrms 50 60Hz of AC sine wave in use to avoid fire or electric shocking please pay attention to all rating values and marks Don t operate if there is any suspected trouble please let the qualified maitainence personnel to check ...

Page 10: ...ng the user to operate it proficiently without taking much time to learn and be familiar with operation of the digital storage oscilloscope In order to speed up the adjustment for measuring the user can press directly the button AUTO and then the instrument shall show the proper waveform and the gear setting Apart from the usability UTD1025CL series of digital storage oscilloscope possess high eff...

Page 11: ...on with 10 probe UTD1025CL Oscilloscope Components Features of UTD1025CL Oscilloscope Oscilloscope USB isolated communication provides safer USB communication Transition from color display to B W display is applicable in the strong outdoor light Smart upgradation function user can upgrade software by local or network Unique waveform recording storage and playback function 5 UTD1000L User Manual No...

Page 12: ...lloscope Please check the new oscilloscope under the following steps if you have one If any serious damage of the package carton or plastic foam protective mat were found please contact with the Uni T distributor or local Uni T representative office immediately As regards to the provided accessory details the above mentioned UTD1025CL Oscilloscopes Accessory in the manual has clearly stated it You...

Page 13: ...stributor that sells the product UNI T shall arrange for the maintenance or replacement The oscilloscope connection is referred to the picture 3 1 1 Oscilloscope channel input port 2 USB connection port 3 Adapter interface used for AC power supply and battery charging 4 Probe compensation signal output port 5 Multimeter input socket 3 circular banana sockets are input ports for voltabe resistance ...

Page 14: ...picture a compensation signal 1kHz 3V output port on the Left side of the apparatus could be used to adjust the probe Panels and keys of oscilloscope are shown in the picture 3 3 Oscilloscope panels and keys 8 Picture Signal Output Connection of the Apparatus Compensation 3 2 Picture 3 3 UTD1025CL Panel ...

Page 15: ...ter the CONFIGURE interface configuration menu Under the multimeter DMM mode press the key to enter into the current measurement menu MEASURE CURSOR R resistance measurement key under the oscilloscope mode press the key to enter COURSOR sursor measurement menu Press the key SHIFT firstly and press it again to enter the MEASURE automatic measurement menu Under the multimeter DMM mode press the key ...

Page 16: ...key to hold screen measurement reading HORIZONTAL HELP key under the oscilloscope mode press the key to enter the HORIZONTAL horizontal setting menu Press key SHIFT and press it again you can open the HELP help information SHIFT key to select functions in combination with other function keys Time base used to change scanning 50s div 10ns div with 1 2 5 system Press S the scanning velocity shall sl...

Page 17: ...cted Customization parameter switch cursor 1 and cursor 2 under the cursor measurement Direction key and OK key OK key used to hide display the present menu bar under normal circumstances under the multimeter DMM mode the keys are used to confirm the correct connection of 10A in A gear current measurement Open trigger main menu and select trigger coupling mode and automatic normal trigger mode by ...

Page 18: ...voltage of which is DC 9V 4A To assure the normal working of apparatus please use the adapter configurated with the product Press power key about 5 seconds the power on LOGO picture shall appear on the oscilloscope and the oscilloscope starts working Picture 4 1 Instruction Picture of Display Interface Apart from the waveform the display will display some detailed information about waveform and os...

Page 19: ...eased collecting waveform data AUTO oscilloscope is under automatic mode and collecting waveform under the non trigger status SCAN oscilloscope collects and displays waveform continuously under the scanning mode 5 Display time reading of trigger point to the center scale 6 Symbols indicate power supply modes as follows 7 Symbols on the screen indicate ground baseline 8 1 symbol indicates the chann...

Page 20: ...ory setting Status after factory setting The following examples introduce how to select function with the oscilloscope menu 1 press function keys on the keyboard the relevant set options to keys shall appear on the bottom of screen select corresponding function with keys F1 F4 after that press key OK to hide function menu UTD1000L User Manual 14 Picuture 4 2 Factory Setting Function Acquire method...

Page 21: ...face configuration setting To switch in signal as the following steps 1 Connect oscilloscope probe with oscilloscope input port and set attenuation multiplying power on the probe as 10 2 Set attenuation coefficient on the probe It shall change the vertical gear multiplying power and the measurement results would reflects correctly the measured signal s amplitude Ways to set the probe attenuation c...

Page 22: ...ture 4 5 Picture 4 4 Probe Multiplying Power Adjusting Picture 4 5 Home Computer Squarewave Signal Probe Compensation Connecting probe with channel for the first time it is required to adjust probe compensation to match probe with the input channel The probe without compensation calibration shall result in measurement error or mistakes To adjust probe compensation under the steps as shown in the a...

Page 23: ...il correct compensation were shown on the screen as the above picture to guard against the electric shock in measuring the high voltage with the probe please make sure that the insulated conductor of probe is intact and do not touch the metal part of the probe when connecting with the high voltage source Press key V mV to adjust vertical gear v div of oscilloscope and to adjust signal to appear on...

Page 24: ...ase gear display for status bar changes correspondingly The horizontal scanning velocity is 10ns 50s div and step in by the way of 1 2 5 Press left right key to adjust horizontal positon of trigger point to observe more pretrigger information When the trigger level symbol is solid press up down key to adjust trigger level posion Horizontal System Trigger Level UTD1000L User Manual 18 Picture 4 7 A...

Page 25: ...plication of automatic setting frequency of the measured signal is required to be bigger than or equivalent to 20Hz duty ratio bigger than 1 UTD1025CL oscilloscope has the creative function of full automatic signal capturing Under the function the oscilloscope shall adjust gear automatically with the changes of input signal and display the waveform in the fittest form without manual intervention T...

Page 26: ...ode the automatic trigger mode shall not be altered 2 Entering into the full automatic setting the following setting shall be compulsory 1 Under the non main time base status it shall switch to main time base status 2 The channel coupling shall switch to AC coupling which is unchangeable 3 To modify vertical scale coefficient of oscilloscope horizontal time base change vertical and horizontal posi...

Page 27: ... count 1 Press key SHIFT character shift shall appear on the right corner of screen 2 Press key UTILITY to show auxiliary function menu 3 Press key F3 to select frequency count function as open To operate under the following steps to stop data collection Press key RUN STOP the oscilloscope shall STOP data collection and trigger status on the upper part of screen shall display STOP Picture 4 10 Sto...

Page 28: ... the right corner of screen 2 Press key ACQUIRE to show backlight menu 3 Press key F1 to opern backlight adjustment box 4 Press right left key to adjust backlight brightness to the proper value 5 Press key to close the backlight adjustment box press key AUTO the backlight adjustment box shall close automatically Notes Picture 4 12 Backlight brightness Adjustment ...

Page 29: ...e 10 groups of bitmap To conduct online help under the following steps 1 Press key SHIFT to show character shift on the right corner of screen 2 Press key HELP to open help function and character H shall appear on the upper part of screen indicating the help function has started 3 Press corresponding menu keys to see detailed help information 4 Reapeat step 1 and 2 to close help function To conduc...

Page 30: ... character shift on the right corner of screen 2 Press key STATUS and the present setting status of oscilloscope shall appear on the screen 3 Reapeat step 1 and 2 to close help status information press key AUTO the status information shall close automatically Notes Picture 4 14 Single Trigger Picture 4 15 Status Information ...

Page 31: ...ce configuration menu 3 Press key F4 the screen shall display informations such as type versionof the present oscilloscope 4 Press key F1 to close system information window There are two ways of automatic parameter measurement showing all parameters or individually set measured parameter to display on screen with maximum of 4 parameter items To measure all parameters automatically under the follow...

Page 32: ...customization parameter menu 2 Press key F2 the screen shall display measurement parameters for choice 3 Press key up down left right to select type of parameter being measured marked by complement color 4 Press key SELECT to confirm the parameter shall display on the screen automatically with set maximum of 4 measuremnt parameters 5 Press key F2 to close customization parameter menu Customization...

Page 33: ...ith 16 Times Smooth Waveform by Average Treatment 1 Press key ACQUIR to show acquisition mode menu 2 Press key F1 to select acquisition mode as average 3 Press key right left to select average frequency as and measurement results shall display times in average To smooth waveform under the following steps ...

Page 34: ...ress key vto select acquisition mode as peak value You can use persistence function to observe dynamic signal continuously 1 Press key DISPLAY to show display mode menu 2 Press key F4 to select 1 second 3 seconds 5 seconds limitless or automatic in a cycle Select Limitless the observed dynamic signal shall persist on the screen and the persistence function shall close if select Automatic Picture 4...

Page 35: ...the DC signal To select AC coupling under the following steps 1 Press key CHANNEL to show channel menu 2 Press key F1 to select channel coupling as AC To display input port waveform in opposite phase under the following steps 1 Press key CHANNEL to show channel menu 2 Press key F4 to select starting opposite phase and the opposite phase waveform shall be displayed on the screen Picture 4 22 AC cou...

Page 36: ...C peak peak value70mV with sine signal containing 1 1V DC bias signal set oscilloscope to display waveform and read reference value Compare UTD1025CL with traditional oscilloscope in terms of operation convenience Connect signal with input channel set channel coupling mode as DC and conduct automatic setting then the screen shall display the waveform shown as in the picture 4 24 Of which Position ...

Page 37: ...ope it is necessary to set it as AC coupling and measure signal s AC parameter set channel as DC manually adjust channel vertical scale coefficient and measure signal s DC parameter by the cursor and get the test value finally as shown in following picture Picture 4 25 AC parameter measurement ...

Page 38: ...ress key DMM DSO the oscilloscope shall switch to multimeter measurement and screen displays multimeter interface 3 4 mm safety banana socket input ports for multimeter COM V A mA One 10A CURRENT DIVIDER UT M07 Instructions 1 Battery indication 2 AUTO automatic setting indication 3 Measurement types indication DC voltage measurement ChapterV Use Multimeter About the Chapter Connect Apparatus Inter...

Page 39: ...SO the oscilloscope shall switch to multimeter measurement and screen displays multimeter interface then you can measure with the multimeter To measure resistance under the following steps 1 Press key R the menu shall display the measurement type resistance 2 Insert blace table pen into COM banana socket input port and red one into banana socket input port 3 Connect the red and blace table pen wit...

Page 40: ...t input port 4 Connect the red and blace table pen with the measured resistor the screen shall display the turn on voltage drop voltage value reading of the diode To conduct continuity test under the following steps 1 Press key R the menu shall display the measurement type resistance 2 Press key F1 the menu shall display the measurement type continuity 3 Insert blace table pen into COM banana sock...

Page 41: ...OM banana socket input port and red one into banana socket input port 4 Connect the red and blace table pen with the capacitor being measured the screen shall display capacitance value reading of the capacitor being measured please use the relative value measurement method in measuring capacitance lower than 5 nF and to improve measurement accuracy measured point resistance is lower than 10 Ω Ω Me...

Page 42: ...the DC voltage value of the measured point as shown in the following picture To measure AC voltage under the following steps 1 Press key V the menu shall display the measurement type DC Voltage 2 Press key F1 the menu shall display the measurement type AC voltage 3 Insert blace table pen into COM banana socket input port and red one into banana socket input port 4 Connect the red and blace table p...

Page 43: ...o mA banana socket input port 3 Connect the red and blace table pen with the measured point and the screen shall display the DC current value of the measured point Picture 5 7 Measurement of DC Current Lower Than 4mA To measure DC current lower than 400mA under the following steps 1 Press key I and the menu shall display measurement type DC current 2 Press key F3 to set range as mA and the measure...

Page 44: ... to set range as A and the measurement unit shall be A 3 Insert 10A CURRENT DIVIDER into current measurement socket and then insert the table pen with the 10A CURRENT DIVIDER 4 Connect the red and blace table pen with the measured point and the screen shall display the DC current value of the measured point Warning in measuring current higher than 400mA the fuse in the machine shall be blown if pl...

Page 45: ...nd the screen shall display the AC current value of the measured point To measure AC current lower than 400mA under the following steps 1 Press key I and the menu shall display measurement type AC current 2 Press key F3 to set range as mA and the measurement unit shall be mA 3 Insert blace table pen into COM banana socket input port and red one into banana 4 Connect the red and blace table pen wit...

Page 46: ...t measurement socket and then insert the table pen with the 10A CURRENT DIVIDER 4 Connect the red and blace table pen with the measured point and the screen shall display the AC current value of the measured point Picture 5 12 Measurement of AC Current higher than 400mA Warning in measuring current higher than 400mA the fuse in the machine shall be blown if plug in port directly without using 10A ...

Page 47: ...relative to the difined reference value The following example introduces how to conduct relative capacitance measurement Firstly to get a reference value 1 Press key R and the menu shall display measurement type resistance 2 Press key F1 the screen shall display measurement type capacitance 3 Insert blace table pen into COM banana socket input port and red one into V banana socket input port 4 Whe...

Page 48: ...key V or R menu display measurement mode automatic 2 Press key F2 to switch measurement mode and enter into manual measurement mode menu display range mode manual 3 Under manual range mode press key F3 once jump a gear upward and press key F3 to the lowest gear when it reaches to the highest gear in a cycle The present range displays above the measurement reading 4 Press key F2 the left upper part...

Page 49: ...on the board and provide some basic examples for using menu and operations You are strongly recommended to read the chapter to know more about the measurement function and system operation methods Channel has independent setting menu press key CHANNEL the system shall display channel operation menu as shown in the following picture Picture 6 1 Channel Menu Setting Channel menu as shown in the foll...

Page 50: ...omponents of the input signal Display DC level of the channel input port being equivalently grounded Display channel waveform Close channel waveform Based on the probe attenuation coefficient select one of them to assure the accuracy of the vertical bias coefficient reading Four types 1 10 100 1000 Normal Waveform display Waveform being displayed in the opposite phase mode ...

Page 51: ...al shift triangle symbol to solid In the channel menu press key F2 to open or close channel waveform display Picture 6 3 Close Waveform Display To cooperate with the probe attenuation coefficient setting it is required to set probe attenuation coefficient correspondingly in the channel operation menu For example the probe attenuation coefficient is 10 1 the probe coefficient in menu shall be corre...

Page 52: ...er and you can see the 6 div pretrigger and delay information Through adjusting waveform horizontal position you can see more pretrigger information and based on the observed pretrigger data to see the waveform status prior to trigger For example if capture the glitch produced during the circuit start time to find the reason by observing and analizing pretrigger data Press key HORIZONTAL to open h...

Page 53: ...se setting Picture 6 5 Screen Display after Window Extension Two display sector in the window extension function as shown in the above picture The upper part displays the original waveform after pressing the key OK to hide menu the sector could be used to shift horizontal extension part left right by adjusting right left key and increase or reduce the selection area by the key s ns The lower half ...

Page 54: ...illoscope integral multiple of and sychoronous to the period of complex waveform being measured For example the signal being measured is a combination wave signal imposed on the channel CH1 Press key HORIZONTAL to open horizontal setting menu and press key left right to adjust holdoff time the trigger holdoff time shall vary with it till the waveform display is stable as shown in the following pic...

Page 55: ...trigger function key TRIGGER TRIGGER Key for trigger function setting menu Trigger level used to set the signal voltage relative to trigger point Press key TRIGGER to open trigger menu as shown in the following picture and press key F1 to switch trigger types Picture 6 7 Trigger Function Menus Four trigger modes edge vedio pluse width and slope different function menu for each trigger and a genera...

Page 56: ...mon trigger options as shown in the following table Picture 6 8 Trigger Setting Common Function menu Trigger Type Trigger source Trigger coupling Trigger modes Set Common set CH1 AC DC High frequency restriction Automatic Normal Instruction Set general trigger options Set CH1 as information source trigger signal Stop DC components contained in the input signal Pass AC and DC components in the inpu...

Page 57: ...the trigger on the trigger threshold of input signal edge Edge trigger setting are as shown in the following table Table 6 4 Pulse width trigge defines trigger time based on the pulse width You can capture abnormal pulse by setting pulse width conditions Picture 6 9 Trigger Edge Trigger Function menu Trigger Type Slope Set Edge Rise Fall Instruction Set trigger type as edge trigger Set trigger on ...

Page 58: ...ons Pulse width time Set Pluse width Positive Negative Instruction Set trigger type as edge trigger Set positive pulse width as the trigger signal Set negative pulse width as the trigger signal Trigger when pulse width equals to setting value Trigger when pulse width is bigger than setting value Trigger when pulse width is smaller than setting value Set pulse width time value under the menu displa...

Page 59: ...o All lines Odd field Even field PAL NTSC Instruction Set trigger type as vedio trigger Vedio signal applicable to Vedio signal applicable to NTSC system PAL system Set vedio horizontal trigger sync Set to trigger sync on the Set to trigger sync on the vedio odd field vedio even field ...

Page 60: ...lope Trigger Select slope trigger and produce trigger when signal rising or falling velocity meets set value conditions Picture 6 12 the First Page of Trigger Slope Trigger Picture 6 13 the Second Page of Trigger Slope Trigger ...

Page 61: ...ween thresholds is bigger than slew rate set value Trigger when signal slope between thresholds is smaller Trigger when signal slope between thresholds equals to slew rate set value than slew rate set value Second page of slope menu Set slew rate value and under the menu display conditions to set pulse width time by left right key Shift low level threshold by up down key Shift high level threshold...

Page 62: ...nabling to display scan baseline on the screen when there is trigger signal the automatic shall switch to trigger scanning and sync with signal when scan waveform is set at the 50ms div or slower time base oscilloscope shall enter into SCAN mode oscilloscope shall collect waveform under normal trigger mode only trigger conditions have been met When there is no trigger signal it stops collecting da...

Page 63: ...example if capturing glitch when the circuit is power on we can find the reason of occurrence of glitch by observing and analyzing pretrigger data Press key ACQUIRE to enter into collection mode menu as shown in the following table Table 6 8 4 Collection Mode Setting Function menu Acquisition Averages appear only the acquisition mode is average Set Normal Peak value Average 2 256 Instruction Set a...

Page 64: ...cts signal to reconstruct waveform with equal time interval under the acquisition mode oscilloscope shall find out the maximum and minimum value of input signal in each acquisition intervals and to display waveform by those values Thus oscilloscope can acquire and display narrow pulse or those narrow pulses will be leaked under the acquisition mode and noise seems louder under the mode oscilloscop...

Page 65: ...ay acquisition Display acquisition in the way of linking points points directly Set mesh display crosshairs mode in the waveform display area full grid Screen waveform updates at normal refresh rate Waveform data on screen updates after holding 1 s Waveform data on screen updates after holding 3 s Waveform data on screen updates after holding 5s Wavefrom data on screen shall hold continuously and ...

Page 66: ...amic changes Y axis represents voltage quantity and X axis time quantity when horizontal time base control is set at 50ms div or slower the apparatus shall enter into slow scan acquisition mode It is recommended to set channel coupling as DC in using slow scan mode toobserve low frequency signal The following introduction shall help you to know more about the powerful automatic measurement functio...

Page 67: ...ing picture Function menu All parameters Select paramter Clear Set ON OFF ON OFF Delete customization parameter displayed on screen Instruction Display measurement value of all parameters Close automatic measurement function Open customization parameter select the needed measured parameter by key SELECT The parameter will appear Close customization parameter selection box selection box to on scree...

Page 68: ...ization parameter selection box indicating the parameter has been customized and will appear on screen concurrently 5 Same to the above operation press up down left right key to select amplitude parameter 6 Press key SELECT to confirm a dot shall appear before the amplitude paramters indicating the parameter has been customized successfully and will appear on screen concurrently 7 Press F2 to clos...

Page 69: ... Parameter Customizations Notes Designated parameters are used for quick parameter measurement There are 19 types of measurement parameters and generally user just need to measure some of them and set as customization parameter instead of all All customization parameters shall appear on screen ...

Page 70: ...id half of amplitude value Peak peak value Vpp voltage value from waveform highest point to the lowest point Amplitude value Vamp voltage value from waveform top to bottom Overshoot ratio of difference between waveform maximum value and top value with amplitude value Preshoot ratio of difference between waveform minimum value and bottom value with amplitude Average value average amplitude of signa...

Page 71: ...width when positive pulse is at 50 amplitude Negative pulse width pulse width when negative pulse is at 50 amplitude Rise delay delay time of A to B rise edge Fall delay delay time of A to B fall edge Positive duty ratio ratio of positive pulse width with period Negative duty ratio ratio of negative pulse width with period Press key RECORD to open waveform record function menu two independent menu...

Page 72: ... Menu Picture 6 23 record file storage menu Table 6 11 Record Menu Function menu Rec Op Play Stop Rec Set Record Instruction Open close waveform record function Play recorded waveform files Stop the waveform files being recorded or played Record waveform ...

Page 73: ...as 1 6 Press key F3 to store recorded waveform and screen shall display saving await saving indicator to disappear the waveform shall be stored into inner memory of the machnine oscilloscope can record at most 300 screens of waveform data and during recording if the following operations are conducted the record shall stop Change horizontal position of signal under stop status Notes Notes Function ...

Page 74: ... to select callback screen will display loading after successful callback indicating Load Success 5 Press key F1 again to select menu function as record 6 Press key F2 to play automatically the callbacked waveform and press key F3 to stop it to manual play with left right key Picture 6 24 Recorded waveform playback To store wavefom under the following steps 1 Press key SHIFT and character shift wi...

Page 75: ...00L User Manual 69 Picture 6 26 Waveform storage menu Function menu Type Location Save Load appear below set menu Set Setting wave 1 20 1 20 Instruction Store sets in oscilloscope present menus Store waveform of channel Select storage position with left right key Save as the former selected operationswith left right key Callback saved apparatus configuration parameters in the set menu ...

Page 76: ...ight key to select storage position as 1 5 Press key F3 to save storage success indicated on screen means save successful and waveform shall be stored into oscilloscope 6 Press key SHIFT and character shift will appear on the right upper corner of screen 7 Press key REF to enter into reference waveform callback set menu Waveform storage Waveform callback Function menu Ref Wave Location Load Close ...

Page 77: ...e There are two cursor measurement types in cursor measurement menu voltage and time In measuring voltage you can use key SELECT on board and up down left right key to adjust cursor position respectively it means to measure if select time you can measure Voltage time measurement method when cursor 1or 2 appear at the same time adjust cursor position with the up down left right key and press key SE...

Page 78: ...rigger Instruction Cursor used to measure time To shift any one of two cursors independently To shift two cursors concurrently maintaining Δt Measurement parameter is time Measurement parameter is frequency To select shift target as channel waveform To select shift target as trigger level To select shift target as measuring cursor and press key SELECT switch cursor to and 1 2 Cursor ...

Page 79: ...ce set menu Function menu Types Modes Function switch Set Amplitude Independent Tracking Channel Trigger Cursor Instruction Cursor used to measure voltage To shift any one of two cursors independently To shift two cursors concurrently maintaining Δt To select shift target as channel waveform To select shift target as trigger level To select shift target as measuring cursor SELECT switch cursor and...

Page 80: ...Skin Menu display Version Set Set Multi lingual Implement system self Adjust Oscilloscope back to factory set status Open close trigger frequency count Delete data information saved in the apparatus such as waveform set etc Classic traditional modern and black white On Off Instruction Instruction Select menu language types Select display style Set awaited close time for menu box 5s 10s 20s and man...

Page 81: ...hall adjust apparatus gear automatically with the changes of input signal and display the waveform in the fittest form without manual intervention Mathematic operation function is used to display operation results of channel waveform FFT To conduct mathematic operation under the following steps 1 Press key MATH to open mathematic operation menu 2 Press key F1 to select FFT operation as open Set as...

Page 82: ...y components in the repeat or single pulse event you can set oscilloscope s acquisition mode as average Suppose YT waveform is repeating continuously oscilloscope takes FFT transformation of the time record with limit length When the period is an integral ampliltude value of YT waveform remains the same at thestart and finish the waveform shall not discontinue UTD1000L User Manual Function menu FF...

Page 83: ...e resolution and worst frequency resolution With the best frequency resolution and worst amplitude resolution Fittest contents for measurement Sine period and narrow band random noise Transient or short pulse signal level varies greatly before after it Mainly used for single frequency signal to find higher order harmonics Transient or short pulse signal level remains the same basically constant am...

Page 84: ... the Nyquist acquisition rate It is used to simplify operation Press key the oscilloscope will adjust vertival bias coefficient and horizontal time base gear based on the amplitude and frequency of waveform automatically and will display stable waveform on screen The system automatic set are as following Automatic Set Table 6 21 UTD1000L User Manual Function menu Acquisition mode Display format s ...

Page 85: ... the battery electricity symbol on the right upper part of screen If symbol indicates battery is depleting charge it Check if channel attenuation coefficient meets with attenuation multiplying power of the probe being used 1 Check if probe is in normal connection with signal connection line 2 Check if signal connection line is in normal connection with BNC channel connector 3 Check if probe is in ...

Page 86: ...ear 2 Press key to complete above settings automatically 1 Generally it is normal to slow down when average frequency is over 32 2 Reduce average frequency 1 It is normal and probably caused by low horizontal time base gear and could be improved by increasing horizontal time base to improve horizontal resolution 2 Probably the display type is vector lines between acquisitions points might result i...

Page 87: ...ire from UNI T company market department or via company website the program upgradation package that type and hardware version is identical to the oscilloscope being upgraded and software version is bigger than that of the present oscilloscope 1 Product type of the program upgradation package shall be in accord with that of the oscilloscope being upgraded 2 Hardware version of the program upgradat...

Page 88: ...d from UNI T market department 3 Connect oscilloscope with computer by USB wire and open the home computer software click button Connect Button in the home computer software interface to realize normal communication between it and oscilloscope After successful connection home computer indicates USB connection successful Picuture 8 1 Connect With Home Computer UTD1000L User Manual ...

Page 89: ...No operation to the oscilloscope at the moment and restart it the program upgradation is completed 1 When there is any mistakes about program upgradation package home computer shall indicate mistakes about upgradation files please check it The oscilloscope shall stop upgradation and the program version shall not change before storage 2 If communication disconnection from home computer power off of...

Page 90: ...es under the stipulate operation temperature If operation temperature shift reaches to or over 5 degrees the system function menu shall be opened to implement the Self correction procedure and except the specification labeled with character Typical all specifications shall be guaranteed Technical UTD1000L User Manual Acquisition mode Acquisition mode Acquisition rate Acquisition Average value Real...

Page 91: ...attenuation coefficient setting Maximum input voltage Time delay in channel typical sin x x Direct current alternative current ground AC GND DC 3 5M 1 2 ΜΩ connect with 20 3PF in parallel 12K 1X 10X 100X 1000X 300V DC AC input impedance peak value 1MΩ 50ps 10ns div 50s divWith 1 2 5 system 50ppm any time interval 1ms Single sampling time interval ppm reading ns 16 average value acquisition time in...

Page 92: ...5MHz 10Hz at the input BNC port 14ns Vertical sensitivity is 5mV div 4 acquisition Vertical sensitivity is 10mV div 20V div 3 acquisition or average value acquisition mode or average value acquisition mode Vertical shift is zero and N 16 4 reading grid mV and select mV div reading grid mV and select mV div 20V div Vertical shft is not zero N 16 reading vertical shift reading Set value mV div to V ...

Page 93: ...Slope trigger Trigger condition Slew rate scope Edge type Trigger level scope Pretrigger ability Holdoff scope 100ns 1 5s 1div positive pulse width negative pulse width 20ns 10s 2divs peak to peak value smaller than bigger than equal to 40pV μs 1 6kV μs Support standard NTSC and PAL Scope horizontally 1 525 NTSC and 1 625 PAL Rise edge fall edge To screen center 10div Normal mode scanning mode pre...

Page 94: ...owable to display cursor in automatic measurement Measurement includes Peak peak value amplitude value maximum value minimum value top value bottom value medium value average value mean square root value overshoot preshoot frequency period rise time fall time positive pulse width negative pulse width duty ratio duty ratio items 19 4 types FFT operation 5 groups of recorded waveform storage 10 grou...

Page 95: ...Hz to 400Hz The accuracy of all measurements is within ofreading number of counts from15 to 25 Add 0 1 specific accuracy for each below15 or above 25 Input impedance 10MΩ 10MΩ valid value Range Range Range Range 400 0mV μA μA 400 0mV 4 000V mA mA 4 000V 40 00V 10A 10A 40 00V 400 0V CURRENT DIVIDER 07 10A use UT M CURRENT DIVIDER 07 10A use UT M 400 0V Accuracy Accuracy Accuracy 1 2 5 character 2 5...

Page 96: ...figuration Display resolution display Fittings Display color Waveform display area Backlighty brightness Backlight intensity Display language types 4 000kΩ 512 0nF 40 00kΩ 5 120uF 400 0kΩ 51 20 F μ 4 000MΩ 100 F μ 40 00MΩ Accuracy Accuracy 3 5inch TFT crystal display An UT M07 10A CURRENT DIVIDER an USB connection wire 320 horizontal RGB vertical picture element 240 Color black white Horizontal 12...

Page 97: ...VDC 4000mA Environment Temperature Cooling method Height Humidity Operation 0 40 Non operation 60 20 Natural cooling Operation 3 000m Non operation 15 000m 35 90 35 40 60 Mechanic specifications Dimension reference data Weight net weight 199mm 118mm 49mm Exclusive of package Inclusive of package 0 9kg 2 3kg IP protection Adjustment interval IP2X Calibration interval is one year recommended ...

Page 98: ...not scrap the transparent LCD protection screen 2 Please power off when clean the instrument with the wet cloth without dripping The soft detergent or clean water could be used to clean the instrument instead of any corrosive chemical detergent which might damage the instrument or probe Before using the apparatus after being powered on again please confirm that it has been dry guarding against the...

Page 99: ...verheat in recharging battery the ambient temperature shall not exceed the allowable value 0 40 Usually it is unnecessary to replace batteries if necessary quantity Five battery signals stipulated in the technical specification it shall be replaced by the qualified personnel and by the lithium ion battery with same specification Notes Notes Replace lithium ion battery It shall not result in any da...

Page 100: ...94 UTD1000L User Manual This user manual may be revised without prior notice ...

Page 101: ... Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon Hong Kong Phone 852 2950 9168 Fax 852 2950 9303 Email info uni trend com Website www uni trend com Manufacturer UNI T Technology Chengdu Co Ltd Add 2F Building D2 No 199 West District Avenue West High tech District Chengdu City Sichuan Province P R China ...

Reviews: