Follow the below procedure to start the cold terminal temperature auto compensation
functions:
1. Press the
pushbutton to turn the Calibrator on.
2. Press
INPUT
to display enter INPUT mode
3. Press
to select J type.
4.
Press and hold
and
together to display 0 or 1 (display 0 is cold terminal
temperature auto compensation, display 1 is the remove cold terminal
temperature auto compensation), release the two buttons.
5. Press
and
to adjust the digit to 0, press
to confirm.
6. Press
pushbutton to turn the Calibrator off.
7.
The setting is finished.
8. At that moment
At simulate TC mode:
The calibrator output thermoelectric force = the corresponding thermoelectric
force of the set temperature minus the corresponding thermoelectric force of
the room temperature.
At Measure TC mode:
The Calibrator display temperature = The corresponding temperature of the
Calibrator input terminal thermoelectric force plus room temperature.
Model UT713: OPERATING MANUAL
12