
Model UT60H: OPERATING MANUAL
39
MAINTENANCE(2)
Warning
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
To replace the Meter’s fuse:
l
Press the
POWER
to turn the Meter off and remove
all connections from the terminals.
l
Remove the fixed rubber feet and 5 screws from the
case bottom.
l
Remove the fuse by gently prying one end loose,
then take out the fuse from its bracket.
l
Install ONLY replacement fuses with the identical
type and specification as follows and make sure the
fuse is fixed firmly in the bracket.
Fuse 1: 0.5A, 250V, fast type fuse,
φ
5x20mm
Fuse 2: 10A, 250V, fast type fuse,
φ
5x20mm
l
Rejoin the case bottom and case top, and install the
5 screws and the rubber feet.
Replacement of the fuses is seldom required. Burning
of a fuse always results from improper operation.
( figure 13)
Screw
C. Replacing the Battery
(see figure 13)
Summary of Contents for UT60H
Page 5: ...Model UT60H OPERATING MANUAL 5 ...
Page 12: ...Functional Buttons ...
Page 13: ...4 5 6 7 8 9 10 11 3 2 1 15 14 13 12 ...
Page 14: ......
Page 16: ......
Page 21: ...Model UT60H OPERATING MANUAL 21 Measurement Operation 7 ...
Page 32: ...Model UT60H OPERATING MANUAL 32 General Specifications ...
Page 34: ...x x ...
Page 35: ......
Page 36: ... ...
Page 37: ......