6. System Setup
This chapater includes the contents as follows.
Page Explaination of <SYSTEM 1>
Page Explaination of < SYSTEM 2>
Page Explaination of <System Information>
6.1 <System> Setup
<System Setup> page is to set some settings that are not related to the specific test item parameters, but are
related to the test scheme of the instrument.
When high voltage output is stop, press [System Setup] key to enter <SYSTEM 1> page.
Figure 6-1 <System Setup 1> Page
Table 6-1 Explaination of <SYSTEM 1> Page
Item
Input Range
Default Value Description
Start
mode
Local
,
PLC
Local
①
Local: Press start key on the front panel to boot-up.
②
PLC: Input START control signal by HANDLER to
boot-up
③
Communication: It’a always valid, the specific
refer to the communication protocol.
Volume
Low
,
middle
,
high
Middle
The volume setting of beeper.
Key sound
ON, OFF
ON
The switch of key sound.
Pass beep
Long and short sound
,
double short sound, OFF
Long
and
short sound
The sound mode of pass beeper.
Fail beep
Long sound, double short
sound, OFF
Long sound
The sound mode of failed beeper.
Start delay (0.1~99.9)s
,
OFF
OFF
The delay time from start to the first test item.