User’s Manual Series
Instruments.uni-trend.com
18
/
40
Data acquisition module of other Instruments are sorted in the same way.
Figure 6-4 UT3232 Channel Number in Acquisition Module of Temperature Tester
6.4 USB
Instrument is equipped with an 8G USB, USB recording function is only valid in <Test>,
<Histogram> and <Curve Chart> page.
Before starting data collection, insert USB into instrument interface and wait the instrument
indentify USB and then press start key, data will be record in file
.
Curve chart and histogram can refer to section 4.5.
It is recommended to use brand USB to avoid the problem of incompatible identification. The
format and capacity of the USB can be referred to FAT, FAT32 and EXFAT, with the maximum
capacity of 128G.
6.5 Operation Guide
This section is to describe how to use multi-channel temperature test to measure DUT. It is
necessary to make full preparation before testing according to the actual situation.
Connect the power line correctly.
Turn on the switch of the instrument, the screen will light up. Check the contact lines of
each temperature are connected well. Starts to data acquisition, if the connection is
wrong then no temperature display on this connection line; if the connection is normal,
temperature will display (environment temperature).