2.Power Inspection
   Please press measurement key and observe display. Battery replacement mark in the left end 
   of display shows low battery voltage and prompt of battery replacement. No mark refers to 
   normal voltage for measurement. (Machine will power off automatically 1 minute after 
   loosening the key.)
3.Selection of measurement mode
   Please toggle measurement selection switch of vibration meter to select measurement 
   acceleration, speed or displacement. Right arrow on display will point to selected 
   measurement unit.
4.Measurement
   Please connect signal wires connected to sensor to input connector of left sensor then absorb 
   sensor with magnetic base to vertical cross section of measured article during usage for the 
   1st time. Instrument can enter into measurement status instantly by pressing measurement 
   key by thumb. Measuring value will be maintained after loosening the key. Measurement also 
   can be continued by pressing measurement key again. Data will be maintained for 1 minute 
   after loosening the key and instrument will power off automatically at the same time. (Note: 
   Please tighten clasp when connecting signal wires and hold instrument and sensor in hands 
   after measurement. It is not allowed to hoist sensor by signal wires to avoid influence to 
   service life of such wires.)

LCD display: 3 1/2-bit display; The maximum display: 1999
Under-voltage display:
Sensor type: Acceleration sensor
Power demand: 1 6F22 9V carbon or alkali battery
Service life: To operate for more than 20 hours continuously
Outline dimension: 185 (H) * 68 (W) * 0 (D) mm (for host)
Weight: About 300g (including battery)

2.Environment Limitation

Indoor usage
Safety standard: GB 4793, 1-2007
Pollution level: Grade 2
Operation temperature & humidity: 0

~40

 (not more than 90%RH)

Storage temperature & humidity: 0

~+40

 (not more than 80%RH)

Atmospheric pressure: 75~106KPa

3.Technical Specification

Accuracy:   5% of 1 word (% of r word quantity); Correction period is 1 year;
Ambient temperature: 0

~40

Ambient humidity: 

80% RH

Temperature coefficient: 0.1*(accuracy)/

VI.TECHNICAL INDICATORS

1. General Specification