IX. Rotary Switch (Fig. 3)

OFF

A

HARM

Phase
Detect

Wh

W

W

3P

Description

Turn off the internal power supply of the product

AC current/voltage measurement

Single-phase power measurement

Three-phase power measurement

Harmonic measurement

Electric energy/electric energy accumulation mode setting

Phase sequence detection

Position

OFF

A/V

W/VA/var

3PW

HARM

Wh/Setting

Phase Detect

X. Button Descriptions

1. SELECT

(1)  A~/V~ position: Short press this button to cyclically select ACA 

→ 

ACV on main display, 

      the default position is ACA.
(2)  W/VA/var position: Short press this button to cyclically select active power → apparent 
      power → reactive power → power factor → phase angel on main display, the default position 
      is active power.
(3) 3PW position: Under 3-phase 3-wire (3P3W, balanced) or 3-phase 4-wire (3P4W) mode, 
      short press this button to cyclically select active power → apparent power → reactive power 
      → power factor → phase angel on main display, the default position is active power. Under 
      3-phase 3-wire (3P3W) mode, short press this button to cyclically select phase-combined 
      active power → P1 active power → P2 active power on main display, the default display is 
      phase-combined active power. Under 3-phase 4-wire (3P4W) mode, short press this button 
      to cyclically select phase-combined active power → phase-combined apparent power → 
      phase-combined reactive power → phase-combined phase angle → P1 active power → P2 
      active power → PC active power on main display, the default display is phase-combined 
      active power.
(4) HARM position: Short press this button to cyclically select voltage harmonic analysis → 
      current harmonic analysis, the default position is current harmonic analysis.
(5) Wh/Setting position: Short press this button to cyclically select single-phases electric 
      energy → electric energy accumulation mode setting, the default position is single-phase 
      electric energy.
(6) Long press this button to turn on/off Bluetooth. 

UT219P User Manual

UT219P User Manual

Fig. 3

2. MAX/MIN

Short press this button to cyclically select real-time value 

 maximum value 

 minimum 

value 

 average value on the main display, and to record the time of capturing maximum 

and minimum values. Long press this button to exit maximum/minimum value measurement. 

The measurement is valid only under such modes including AC voltage, AC current, single-

phase power, balanced three-phase power, etc.

3. RANGE

When this button is short pressed in auto range mode, the product enters manual mode, the 

LCD shows “RANGE: MANUAL”, and the display is in current range. Press this button 

continuously to select AC current ranges cyclically. When this button is long pressed, the 

product exits manual range and enters auto range, “RANGE: MANUAL” does not appear on 

the LCD.

4. HOLD/ 

Under measurement interface (except unbalanced three phase, electric energy and phase 

sequence), short press this button to enter or exit data hold. When entering data hold mode, 

the LCD shows “HOLD”; when exiting data hold mode, the LCD does not show “HOLD”. 

Under the interface of switching wire connection, three-phase power enters its corresponding 

wire-connection measurement interface when this button is short pressed.

Under electric energy mode, short press this button to start or stop accumulative counting.

Long press this button to turn on/off backlight. The backlight turns off automatically after it is 

turned on for one minute.

5.     /3P3W

For 3PW position, short press this button to select measurement mode of 3P3W balanced 

load and unbalanced load.

Under harmonic measurement mode, short press to adjust down the order of harmonic.

6.     /3P4W

For 3PW position, short press this button to select measurement mode of 3P4W balanced 

load and unbalanced load.

Under harmonic measurement mode, short press to adjust up the order of harmonic.

Note: For the operating instructions of button of setting electric energy accumulation mode, 

please refer to the operating instructions of measurement in electric energy position.

XI. Operating Instructions

Please check the three pieces of AAA batteries before measurement, if the battery voltage is 
low, the symbol “        ” will appears on the LCD, then please replace the battery in time. Note 
the symbol “      ” around the terminal, this symbol warns that the measured voltage cannot 
exceed the specified voltage.

Summary of Contents for UT219P

Page 1: ...P N 110401111507X UT219P AC Digital Power Clamp Meter User Manual...

Page 2: ...end guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misuse...

Page 3: ...1000V AC 4 1000V AC RMS voltage measurement 15Hz 1000Hz frequency response range 5 1000A AC RMS current measurement 15Hz 1000Hz frequency response range 6 Ability to measure and display RMS and peak...

Page 4: ...uracy Remove the battery if the product is not used for a long time 10 Do not alter the internal wiring to avoid product damage and safety hazard 11 Do not keep or use the product in environments with...

Page 5: ...ctor 4 3P3W 3 phase 3 wire measurement mode 5 3P4W 3 phase 4 wire measurement mode 6 UNBALANCED Unbalanced measurement mode 7 MAX Maximum value measurement 8 MIN Minimum value measurement 9 AVG Averag...

Page 6: ...ximum value minimum value average value on the main display and to record the time of capturing maximum and minimum values Long press this button to exit maximum minimum value measurement The measurem...

Page 7: ...ip already with V terminal and black connected with black alligator clip already with COM 2 Set the rotary switch to A V press SELECT to select AC voltage measurement and connect the alligator clips w...

Page 8: ...lamp jaws are correct L1 R A L2 S B L3 T C Red Black Power supply AC 3 phase 3 wire power measurement 3P3W balance Load 4 3 phase power measurement 4 1 3 phase 3 wire 3P3W balanced mode Fig 8 Fig 8 UT...

Page 9: ...voltage Auxiliary display 3 AC current Short press Short press Short press Short press Short press 4 3 3 phase 3 wire 3P3W unbalance mode Fig 10 L1 R A L2 S B L3 T C Red Black Power supply AC 3 phase...

Page 10: ...alligator clip already with V terminal and black connected with black alligator clip already with COM 2 Set the rotary switch to 3PW press to select 3P4W unbalance mode then the clamp enters P1 wire c...

Page 11: ...ower Auxiliary display 1 P1 prompt Auxiliary display 2 Voltage Auxiliary display 3 Current Main display Second active power Auxiliary display 1 P2 prompt Auxiliary display 2 Voltage Auxiliary display...

Page 12: ...at the center of the clamp jaws Error of 1 0 of reading will be produced if the measured conductor is not placed at the center of the clamp jaws If zero crossing of harmonic does not meet the zero cr...

Page 13: ...tronic meter Press Hold to start accumulation when the LED flashes for one time press HOLD again to stop accumulation when the LED flashes again as shown below Electronic Meter LED flashes for one tim...

Page 14: ...ace of wire connection Fig 13 within 10 seconds connect test lead from V terminal with L3 live wire and connect test lead from COM terminal with L2 live wire then three kinds of testing results will b...

Page 15: ...al error of temperature coefficient is 0 1 specified accuracy C To ensure measurement accuracy please place the measured conductor at the center of the clamp jaws Error of 1 0 of reading will be produ...

Page 16: ...angle error 3 5 Phase angle 15Hz 40Hz 70Hz 1000Hz 40Hz 70Hz 1000A AC 1000Vrms 0 1 3 5 5 Range Resolution Accuracy Overload protection Remark Min measured voltage 30V Max measured current 10A 180 adva...

Page 17: ...ment result viewing button control and other operations can be achieved through UNI T Smart Measurement APP 3 2 UNI T Smart Measurement APP has multiple functions including Bluetooth communication dat...

Reviews: